High Power Laser and Particle Beams, Volume. 33, Issue 4, 046002(2021)

Personal dosimeter calibration based on minitype reference radiation

Yikun Qian1...2, Peng Feng2, Yixin Liu1, Yadi Zhu1, Yuchen Huang1,2, Song Zhang1,2, Peng He2, Biao Wei2 and Benjiang Mao1,* |Show fewer author(s)
Author Affiliations
  • 1China Academy of Engineering Physics, Mianyang 621900, China
  • 2College of Optoelectronics Engineering, Chongqing University, Chongqing 400044, China
  • show less
    Figures & Tables(11)
    [in Chinese]
    [in Chinese]
    [in Chinese]
    [in Chinese]
    [in Chinese]
    [in Chinese]
    [in Chinese]
    • Table 1. [in Chinese]

      View table
      View in Article

      Table 1. [in Chinese]

      thickness of stage/mmdose rate at the point of test/(mSv/h)impact rate/%error/%
      51.3831.140.11
      101.3952.000.11
      151.4042.700.11
      201.4123.270.11
      251.4193.780.11
      301.4244.200.11
      351.4304.570.09
      401.4344.900.09
    • Table 2. [in Chinese]

      View table
      View in Article

      Table 2. [in Chinese]

      size of dosimeter/(mm×mm×mm)dose rate at the point of test/(mSv/h)impact rate/%error/%
      10×3×81.4271.620.03
      8×3×81.4191.080.04
      8×3×61.4110.460.04
      8×3×41.397−0.470.05
      8×2×41.4140.700.05
      8×1×41.4201.110.05
    • Table 3. [in Chinese]

      View table
      View in Article

      Table 3. [in Chinese]

      number of dosimeter/mmdose rate at the point of test/(mSv/h)impact rate/%error/%
      11.4000.100.10
      21.4010.150.09
      41.4020.240.10
      81.4070.560.11
    • Table 4. [in Chinese]

      View table
      View in Article

      Table 4. [in Chinese]

      uncertainty factorreference conditionerror value/%uncertainty component/%
      MRR box60 cm×60 cm×60 cm0.080.05
      thickness change of stage20 mm3.271.89
      size change of dosimeters10 mm×3 mm×8 mm1.620.94
      number change of dosimeters80.560.32
      combined standard uncertainty3.20
    Tools

    Get Citation

    Copy Citation Text

    Yikun Qian, Peng Feng, Yixin Liu, Yadi Zhu, Yuchen Huang, Song Zhang, Peng He, Biao Wei, Benjiang Mao. Personal dosimeter calibration based on minitype reference radiation[J]. High Power Laser and Particle Beams, 2021, 33(4): 046002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Nuclear Science and Engineering

    Received: Oct. 11, 2020

    Accepted: --

    Published Online: Jun. 24, 2021

    The Author Email: Mao Benjiang (gena@vip.sina.com)

    DOI:10.11884/HPLPB202133.200280

    Topics