Acta Optica Sinica, Volume. 30, Issue 10, 2923(2010)

A New Calibration Method of Systematic Errors in Phase-Shifting Point Diffraction Interferometer

Liu Ke* and Li Yanqiu
Author Affiliations
  • [in Chinese]
  • show less

    Because the phase-shifting point diffraction interferometer (PS/PDI) uses lateral shearing of test and reference waves to generate carrier frequency fringe pattern, significant systematic measurement errors are consequentialty introduced. A new calibration method of PS/PDI systematic errors, named rotating grating method, is proposed. Based on the PS/PDI measurement results obtained with two gratings in orthogonal directions, the proposed method utilizes the orthogonal and symmetric property of Zernike polynomials in unit circle to calibrate the systematic errors of PS/PDI. The calibration experiment of visible light PS/PDI is performed to verify the method. The result shows that the geometric coma of the system is -0.116 λ which is consistent with the theoretical value of -0.133 λ, which proves the validity of the method.

    Tools

    Get Citation

    Copy Citation Text

    Liu Ke, Li Yanqiu. A New Calibration Method of Systematic Errors in Phase-Shifting Point Diffraction Interferometer[J]. Acta Optica Sinica, 2010, 30(10): 2923

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Nov. 24, 2009

    Accepted: --

    Published Online: Oct. 24, 2012

    The Author Email: Ke Liu (liukess2008@gmail.com)