    Figures & Tables(4)
    Schematic block diagram of calibration device
    Schematic diagram of system for FTIR
    Schematic block diagram of system for signal acquisition and processing
    • Table 1. Measurement results of transmittance of neutral density filter in 2 µm~14 µm band

      View table
      View in Article

      Table 1. Measurement results of transmittance of neutral density filter in 2 µm~14 µm band

      测量次数测量装置
      红外滤光器透射比测量装置/%红外光学系统光谱透射比校准装置/%
      156.8656.71
      256.6857.44
      356.8757.40
      456.6657.81
      556.7957.02
      656.7657.32
      平均值56.7757.27
      相对标准偏差0.301.00