Infrared and Laser Engineering, Volume. 49, Issue 4, 0413006(2020)

Integrated imaging system of eye axis and anterior segment based on SS-OCT

Huaiyu Cai1... Weiqian Zhang1, Xiaoyan Han1, Yi Wang1, Wenguang Chen2 and Xiaodong Chen1 |Show fewer author(s)
Author Affiliations
  • 1Key Laboratory of Opto-Electronics Information Technology of Ministry of Education, College of Precision Instrument and Opto-Electronic Engineering, Tianjin University, Tianjin 300072, China
  • 2Shanghai Medi Works Precision Instrument Incorporated Company, Shanghai 200237, China
  • show less
    Figures & Tables(10)
    Fundamental principle and the display of SS-OCT system
    Flow chart of SS-OCT system data processing
    Schematic diagram of data acquisition sequence
    Sample's interference spectra before and after spectral reshaping
    K-trigger signal and signal before and after k-resampling
    Flow chart of image enhancement
    Original image and the enhanced results
    Full-scale eye model measurement result
    • Table 1. Comparison between the measurement of the parameters of each layer of the eye axis and the actual sample size

      View table
      View in Article

      Table 1. Comparison between the measurement of the parameters of each layer of the eye axis and the actual sample size

      MeasurementCCT/mmAD/mmLT/mmVT/mmAL/mm
      Sample size0.333.064.0618.9726.42
      Result without calibration0.353.084.0719.1026.60
      Result after calibration0.323.074.0719.0026.46
      Error value0.010.010.010.030.04
    • Table 2. Comparison results with traditional segmented measurement system

      View table
      View in Article

      Table 2. Comparison results with traditional segmented measurement system

      SourceMethodMeasurement error/mm Imaging dimensionNumber of layers Time/s
      Tao[13]Reference arm with variable optical path0.452D50.9
      Shanghai Jiao Tong University[14]Dual band dual focus OCT0.772D/3D510
      Marco[28]Reference arm with variable optical path0.022D53.4-4
      This systemLarge-scale measuring system0.042D50.45
    Tools

    Get Citation

    Copy Citation Text

    Huaiyu Cai, Weiqian Zhang, Xiaoyan Han, Yi Wang, Wenguang Chen, Xiaodong Chen. Integrated imaging system of eye axis and anterior segment based on SS-OCT[J]. Infrared and Laser Engineering, 2020, 49(4): 0413006

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Dec. 11, 2019

    Accepted: --

    Published Online: May. 27, 2020

    The Author Email:

    DOI:10.3788/IRLA202049.0413006