Chinese Journal of Lasers, Volume. 35, Issue 9, 1429(2008)

Ophthalmic Imaging by Optical Coherence Tomography

Shi Guohua1,2、*, Ding Zhihua3, Dai Yun1, Rao Xuejun1, and Zhang Yudong1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • 3[in Chinese]
  • show less

    Single-mode fiber-designed optical coherence tomography (OCT) system is developed and applied in ophthalmic imaging. Images with high signal-to-noise ratio, high resolution and large image depth, are achieved by the developed system. The calibrated axial resolution and transversal resolution are 9 μm and 10 μm, respectively, and the maximum depth range is 3.4 mm. Comparison between experimental results and Zeiss OCT Ⅲ are presented. The results show that the proposes OCT system can image not only the layers of retina , but also the layers and vas of choroid which cannot be obtained by Zeiss OCT Ⅲ.

    Tools

    Get Citation

    Copy Citation Text

    Shi Guohua, Ding Zhihua, Dai Yun, Rao Xuejun, Zhang Yudong. Ophthalmic Imaging by Optical Coherence Tomography[J]. Chinese Journal of Lasers, 2008, 35(9): 1429

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: biomedical photonics and laser medicine

    Received: Oct. 11, 2007

    Accepted: --

    Published Online: Sep. 9, 2008

    The Author Email: Guohua Shi (shighua19810410@sina.com.cn)

    DOI:

    Topics