Chinese Journal of Lasers, Volume. 35, Issue 9, 1429(2008)
Ophthalmic Imaging by Optical Coherence Tomography
Single-mode fiber-designed optical coherence tomography (OCT) system is developed and applied in ophthalmic imaging. Images with high signal-to-noise ratio, high resolution and large image depth, are achieved by the developed system. The calibrated axial resolution and transversal resolution are 9 μm and 10 μm, respectively, and the maximum depth range is 3.4 mm. Comparison between experimental results and Zeiss OCT Ⅲ are presented. The results show that the proposes OCT system can image not only the layers of retina , but also the layers and vas of choroid which cannot be obtained by Zeiss OCT Ⅲ.
Get Citation
Copy Citation Text
Shi Guohua, Ding Zhihua, Dai Yun, Rao Xuejun, Zhang Yudong. Ophthalmic Imaging by Optical Coherence Tomography[J]. Chinese Journal of Lasers, 2008, 35(9): 1429