Reseach on calibration method of double-channel thermal-infrared standard radiometer

Xie Chenyu1,2、*, Li Jianjun1, Pang Weiwei1, Xia Maopeng1, and Zheng Xiaobing1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    Blackbody radiation source as a primary standard, its measurement accuracy determines the uncertainty level of the whole traceability link. To improve the measurement accuracy of blackbodies′ radiance, a double-channel thermal-infrared standard radiometer was developed. According to the principle of the instrument, calibration coefficients of the standard radiometer were obtained by using a high accuracy water-bath blackbody. Experimental results suggest that the radiometer′s instability is less than 0.3% within one hour, uncertainty of fitting coefficient is less than 1.23%, and combined uncertainty is less than 0.39%, 1.3% in 5 μm and 10 μm channel respectively. This radiometer provides new means for tracing blackbody radiance directly to radiant standard in lab.

    Tools

    Get Citation

    Copy Citation Text

    Xie Chenyu, Li Jianjun, Pang Weiwei, Xia Maopeng, Zheng Xiaobing. Reseach on calibration method of double-channel thermal-infrared standard radiometer[J]. Infrared and Laser Engineering, 2018, 47(4): 404002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: 红外技术及应用

    Received: Nov. 10, 2017

    Accepted: Dec. 20, 2017

    Published Online: Sep. 19, 2018

    The Author Email: Chenyu Xie (xcy1993@mail.ustc.edu.cn)

    DOI:10.3788/irla201847.0404002