Reseach on calibration method of double-channel thermal-infrared standard radiometer
Blackbody radiation source as a primary standard, its measurement accuracy determines the uncertainty level of the whole traceability link. To improve the measurement accuracy of blackbodies′ radiance, a double-channel thermal-infrared standard radiometer was developed. According to the principle of the instrument, calibration coefficients of the standard radiometer were obtained by using a high accuracy water-bath blackbody. Experimental results suggest that the radiometer′s instability is less than 0.3% within one hour, uncertainty of fitting coefficient is less than 1.23%, and combined uncertainty is less than 0.39%, 1.3% in 5 μm and 10 μm channel respectively. This radiometer provides new means for tracing blackbody radiance directly to radiant standard in lab.
Get Citation
Copy Citation Text
Xie Chenyu, Li Jianjun, Pang Weiwei, Xia Maopeng, Zheng Xiaobing. Reseach on calibration method of double-channel thermal-infrared standard radiometer[J]. Infrared and Laser Engineering, 2018, 47(4): 404002