Journal of Infrared and Millimeter Waves, Volume. 44, Issue 3, 425(2025)

Analysis of the effect of vibration on the homogeneity of the kinematic mechanism of an angular mirror translational interferometer

Xue-Rong XU1,2, Yi-Tian PENG1、**, Ming-Jian GU2、*, and Teng-teng JIANG3
Author Affiliations
  • 1College of Mechanical Engineering, Donghua University, Shanghai 201620, China
  • 2Shanghai Institute of Technical Physcics Chinese Academy of Sciences, Shanghai 200083, China
  • 3State Key Laboratory of Infrared Physics,Shanghai Institute of Technical Physics,Chinese Academy of Sciences,Shanghai 200083,China
  • show less

    The effect of external vibration on the velocity uniformity of the moving mechanism of the angular mirror translational Fourier transform interferometer (hereinafter referred to as interferometer) can be quantitatively analysed by the interferometer optical range difference velocity stability. The article proposes a more comprehensive method of analysing the optical range difference velocity uniformity for the reliability of the interferometer kinematic mechanism under the influence of on-orbit microvibration in the process of space spectroscopy detection. The method incorporates the structural response of the interferometer caused by external excitation into the stability analysis as one of the influencing factors, so as to reflect the reliability of the interferometer in orbit more realistically, and judge the microvibration criticality that the interferometer can withstand more accurately. At the same time, an optical surface model of the interferometer is established to further theoretically characterise the effect of microvibration on the homogeneity of the interferometric mechanism. The method discussed in the article provides a way of thinking for the judgement of the reliability of the mechanism movement under the external excitation perturbation as well as the research on the optimisation of the mechanism control.

    Keywords

    Introduction

    The interferometric Fourier transform spectral detection technology has many advantages such as large detection luminous flux, multiple spectral channels, high spectral resolution, etc. It has gradually become a hot spot for the development of spectral detection technology at home and abroad1-3 . The purpose of spectral detection is to obtain the spectral information of the measured beam. Spectral detection firstly modulates the measured beam in time or space, and obtains the interference signal of the measured beam; and then carries out Fourier inversion on the interference signal, obtains the spectral information of the measured beam, and completes the task of spectral detection. In the process of spectral detection, the interferometric signal is obtained by the interferometer (in this paper, it refers to the angular mirror translational interferometer), and the interferometric signal directly determines the quality of the spectral map, so the interferometer technology is the key to the spectral detection4 . In order to obtain high-quality interferometric signals, it is necessary to ensure the precision movement of the linear mechanism of the interferometer itself to achieve the uniform scanning modulation of the interferometric optical range difference. And the high-precision spectral detection puts forward extremely high requirements on the uniformity of the interferometer optical range difference speed. However, after orbiting, the micro-vibration of the flywheel, scanning mechanism, mechanical cooler and other equipments on the satellite will affect the speed uniformity of the interferometer.

    Generally speaking, the velocity uniformity of the interferometer is mainly guaranteed by the mechanism control system, but the actual engineering often encounters the situation that the control system's guarantee of uniformity is not clear due to the immaturity of the control algorithm. In order to clarify the impact of microvibration on the interferometer under such circumstances, and to determine the critical value of microvibration that the interferometer can withstand, this paper proposes a quantitative analysis and evaluation method.

    1 Homogeneity of interferometer optical range difference velocity

    The velocity uniformity of an interferometer's motion mechanism can be quantitatively determined by the stability of the interferometric optical range difference velocity. In order to satisfy high-precision spectral detection tasks, interferometers need to ensure a sufficiently high stability of the optical range difference velocity, which has to be realised by the precision motion of the kinematic mechanism5 .

    1.1 Interferometer optical range difference

    The angular mirror translation interferometer achieves the scanning of the optical range difference by the translation of the angular mirror (cubecorner)driven by the voice coil motor and the fixation of the fixed mirror. The interferometer optical path is shown in Figure 1.

    Interferometer optical path difference (OPD)

    Figure 1.Interferometer optical path difference (OPD)

    In Fig. 1, the measured collimated beam is divided into a reflected beam A on the beam splitter mirror and Fig. 1, the measured collimated beam is divided into a reflected beam A and a transmitted beam B on the beam splitter mirror, the A beam is reflected back through the angular mirror CC1 and reflected through the beam splitter mirror and reaches the detector; while the B beam is reflected back through the angular mirror CC2 and then reflected through the beam splitter mirror and reaches the detector as well. In this way, beams A and B produce interferogram signals on the detector. Definition of A, B beams of light equal position, that is, the figure CC1 and CC2 location for the interferometer Zero Path Difference position (ZPD).

    When the voice coil motor drives the angular mirror CC2 to move a distance L, relative to the ZPD position, the B beam travels 2 mechanical path differences L, so at this time the Optical Path Difference (OPD)between beams A and B is 2L.

    Thus the optical range difference D ofthe angular mirror translational interferometer is6-7

    D=2L(t) .

    1.2 Interfering optical range difference velocity stability

    The optical range difference velocityv=ddtD(t) , according to Eq. (1).

    v=2ddtL(t) .(2)

    From Eq. (2), it can be seen that the optical differential velocity depends on the velocity ddtL(t). Uniform scanning modulation of the interferometer optical range difference requires that the speed v of the optical range difference be uniform.

    The uniformity of the optical range difference velocity is generally quantitatively described by the optical range difference velocity stability. Firstly, the instability of the optical range difference velocity is defined as α. α is the ratio of the standard deviation to the mean value of all optical range difference velocity samples taken by the interferometer angular mirror in the effective range of the optical range, and it is generally required that α <1%8-9 ; the stability of the optical range difference velocity is defined as (1-α), and it is generally required that (1)is >99%.

    2 Interferometer optical surface rigid body motion

    For analysis using finite elements, the finite element model of an optical element is defined as a shell or solid cell, and the optical surface is characterised by a number of grid nodes,with the average rigid-body motions of the surface (translational displacements TxTyTz and rotational displacement RxRyRz) is characterised by an area-weighted average motion3. For a nodeixi,yi,zi its rigid-body displacements dxidyi and dzi are expressed as

    dxi=Tx+ziRy-yiRzdyi=Ty-ziRz+xiRzdzi=Tz+yiRz-xiRy .

    According to the principle of least squares best fit, the surface displacements obtained from the finite element analysis are post-processed using Matlab programming, then the actual optical element surface node displacements DxiDyiDzi (vector-height displacement)and the nodal rigid body displacement dxidyi and dzi The squared difference between the actual optical element surface node displacement (vector height displacement)and the node rigid body displacementE is

    E=iωi[Dxi-dxi2+Dyi-dyi2+Dxi-dzi2] .

    Eq:ωi is the weighting factor.

    The best fit for the rigid body motion is obtained by substituting equation (3) into equation (4), taking partial derivatives for each term and making its derivative equal to dry zero. The resulting system of simultaneous equations for the mean rigid body motion is

    iωiDxi=iωiTx+iωiziRy-iωiyiRziωiDyi=iωiTy+iωixiRz-iωiziRxiωiDsi=iωiTz+iωiyiRx-iωixiRyiωiziDyi-yiDsi=iωiziTy+xiRz-iωiyiTz-xiRy-iωiRxzi2+yi2iωiziDxi-xiDsi=iωiziTx-yiRz-iωixiTz+yiRx+iωiRyzi2+xi2iωiyiDxi-xiDyi=iωiyiTx+ziRy-iωixiTy+ziRx-iωiRzyi2+xi2 .

    The optical model of the angular mirror translational interferometer is shown in Fig. 2. An uptake is applied to the moving mirror fixed mirror and beam splitter along the three axes and in the direction of rotation around each axis to obtain the sensitivity coefficients of the system's optic-axis error, each of which is the offset of the image divided by the input value of the uptake. For this optoelectronic system, ignoring the terms with small sensitivity coefficients, the equation of motion of the image in the focal plane can be derived as

    Δx=φM1,TxTM1,x+φM1,RyRM1,y+φM2,TxTM2,x+    φM2,RyRM2,y+φIP,TxTIP,xΔy=φM1,TyTM1,y+φM1,RxRM1,x+φM2,TyTM2,y+     φM2,RxRM2,x+φHP,TyTHP,y

    Finite element model of the interferometer

    Figure 2.Finite element model of the interferometer

    Eq:φ is the sensitivity coefficient;TR denote the translational and rotational displacements, respectively; the subscripts M1, M2, and IP correspond to the moving mirror fixed mirror and beam splitter, respectively;Δx and Δy are the image motion inx andy direction of the image motion, respectively.

    3 Methods of analysing the effects of microvibration

    The work in this paper is aimed at identifying the critical value of microvibration that the interferometer can withstand when the mechanism control system's guarantee of the uniformity of the speed of the moving mirror's linear motion is not clear. Since the influence of the control system is not clear, the analysis process assumes that it has no influence on the external environment. It is assumed that the linear travel of the moving mirror at any transient moment under microvibration is a linear superposition of the pendulum angle of this control system and the pendulum angle due to microvibration. Such a linear travel of the moving mirror of the interferometer is then used to derive results on the homogeneity of the differential velocity of the optical path, and to quantify the effect of the microvibration on the reliability of the interferometer under a given mechanism control system.

    This method of analysis incorporates the microvibration-induced interferometer response into the analysis of the mechanism's kinematic stability, which ensures more realistic reliability analysis results.

    3.1 Underlying assumptions

    Microvibration excitation directly affects interferometer travel Dt and motion velocity ddtD(t). It is assumed that under microvibration conditions, the motion travel Dt is the motion stroke controlled by the mechanism control system Dct and the motion travel caused by microvibration Ddt The linear superposition of, i.e.

    Dt=Dct+Ddt .

    It should be noted that here the movement stroke of the mechanism control system Dct does not include velocity uniformity then interferometer velocity uniformity is guaranteed.

    Then the speed of motion of the moving mirror of the interferometer

    ddtDt=2ddtLct+2ddtLdt         .

    At this point the optical range difference of the interferometer is

    D=2Lct+2Ldt .

    It can be seen that the differential optical velocity under the influence of microvibration is a function of 2 variables.

    3.2 Kinematic properties of translational angular mirrors

    With the above mechanism control system, the interferometer moving mirror movement distance Dct and speed ddtDct are derived. According to equation (2)there are

    vt=Lt+c

    where c is a constant. Definition t = 0, the interferometer angular mirror is located in the optical zero position (ZPD), at this time the interferometer angular mirror movement time t = 0, so c = 0. Therefore, the stroke of the interferometer angular mirror under the mechanism control system is Lt Therefore, the travel of the interferometer mirror under the mechanism control system is

    vt=Lt .

    The microvibration-induced velocity response v and velocity response dv)of the interferometer's moving mirror angular mirror travel can be obtained by discrete time-domain response solving with a finite element model of the interferometer structure.

    4 Analysis of microvibration effects in angular mirror translational interferometers

    The angular mirror translational interferometer, shown in Figure 2, is connected to the satellite platform by means of mounting lugs. The platform is equipped with a flywheel and carries multiple scanning mechanisms and chiller loads. The platform vibration excitation frequencies are concentrated at 90 Hz, 70 Hz, 50 Hz and 30 Hz, and the micro-vibration is excited in the three main-axis directions simultaneously on the interferometer mounting lugs. In order to clarify the effect of the platform perturbations on the operation of the interferometer, the velocity uniformity of the interferometer under the above excitations is analysed using the method described in this paper.

    The interferometer has two modes of operation and the associated parameters are shown in Table 1.

    Firstly, the travelling distance and speed of the motion under a certain mechanism control system are solved. According to the Rayleigh criterion, the spectral resolution limit is 12M which is determined by the maximum optical range difference generated by the moving mirror movement. Therefore, the larger the optical range difference, the higher the spectral resolution of the instrument.

    Then using the interferometer structure finite element model (Figure 3), the time domain response is solved to obtain the response of the pendulum arm angle and angular velocity caused by microvibration. According to the frequency characteristics of the platform disturbance source, the microvibration conditions are selected as 12 working conditions with frequencies of 90 Hz, 70 Hz, 50 Hz, 30 Hz, and acceleration amplitudes of 50×10-3gn, 30×10-3gn, and 10×10-3gn combinations, and the simultaneous excitation of three spindle directions is considered.

    • Table 2. Optical range difference velocity stability under microvibration conditions

      Table 2. Optical range difference velocity stability under microvibration conditions

      Microvibration frequency / HzOptical range difference velocity stability/%
      50×10–3gn30×10–3gn10×10–3gn
      9098.5498.499.3
      7098.698.7699.39
      5098.8699.3299.57
      3099.3499.4899.81
    • Table 1. Parameters of the interferometer

      Table 1. Parameters of the interferometer

      Interferometer characteristic termsWorking mode
      Maximum optical range difference during motion Dmax /mm5
      Required differential optical velocity v / (mm/s)0.2
      Spectral Resolution Mσ (cm-10.5

    Model of the interferometer

    Figure 3.Model of the interferometer

    Thus, the mean value of the optical range difference velocity is E=i=1Nv(ti)N The standard deviation of the optical differential velocity is σ=i=1N[vti-E]2N The stability of the optical differential velocity under the above microvibration conditions is shown in Table 2.

    Finally, the obtained control system and the results of the microvibration-induced angle-mirror travel analysis are substituted into Eq. (2)to obtain the time-domain values of the optical differential velocities of the interferometer angle-mirror motion under microvibration conditionsvtii =1,2,…, N, where N is the total number of moments in the finite element time domain discretisation. The time-domain transient process of the interferometer mirror travelling is obtained by solving the time-domain curve as in Fig. 4. This overview can be approximated as a linear function and is represented as a target velocity value in Fig. 4.

    Analysis results in discrete time sequence

    Figure 4.Analysis results in discrete time sequence

    The time-domain curve contains information about the value of the mirror travelling caused by the whole time-domain process of the microvibration action. Fig. 5 is the schematic diagram of the mechanical experiment. The transient process of the velocity motion of the moving mirror is shown in Fig. 4 under the conditions of 50×10-3gn,30×10-3 gn and 10×10-3gn with frequencies of 90 Hz.

    Instrument structure of the interferometer based on the angular mittor translation system

    Figure 5.Instrument structure of the interferometer based on the angular mittor translation system

    Analysing Table 2 the following conclusions can be obtained, under a certain control system:

    1)The smaller the magnitude of the vibration acceleration, the higher the stability for a given microvibration frequency.

    (2)For a certain magnitude of microvibration acceleration, the lower the vibration frequency, the higher the stability.

    (3)Overall, when the microvibration amplitude magnitude is in the order of 10×10-3gn, the interferometer optical range difference velocity stability (1-α)at all frequencies meets the basic requirement of >99%.

    Desktop micro-vibration tests were carried out on the flat-acting angular mirror mechanism, and the test results are shown below. A comparison of the test programmes is shown in Fig. 5.

    As shown in Table 2, at the same amplitude, the velocity uniformity becomes worse as the frequency increases, so the Fourier transform calculation is performed at 90 Hz when the velocity uniformity is greater than 99% and the vibration conditions are the worst.The result, shown in Figure 6, meets the design criteria.

    Plot of test results of the planar angular mirror mechanism programme

    Figure 6.Plot of test results of the planar angular mirror mechanism programme

    5 Conclusions

    To address the problem of the interference of in-orbit microvibration on the precision motion of spectral detection interferometer mechanisms, a more comprehensive method of analysing the interferometer optical range difference velocity uniformity is proposed. The method incorporates the microvibration-induced structural response of the interferometer as an influencing factor into the analysis of interferometer stability, which can reflect the reliability of the interferometer on-orbit working state more realistically than the usual method of equating the influence of external excitation to a certain influencing factor10-14 . This method allows to quickly determine the critical values of the interferometer's microvibration magnitude in conditions where the organisation's control algorithms for external environmental excitations are not well defined. Considering that the control algorithm is usually an improvement of the external environment, this critical value has a certain safety margin.Therefore, this method has some application value in solving practical engineering problems.

    [1] XING T, GONG H X. Space borne fourier transform spectrometer for atmospheric sounding[J]. Remote Sensing Technology and Application, 14, 5-10(1999).

    [4] Giroux Jacques, Moreau Louis, Girard Guillaume et al. Technological evolutions on the FTS Instrument For Follow-On Mission to SCISAT atmospheric chemistry experiment[J]. Proceedings of SPIE, 7826, 7826-1A(2010).

    [9] XU Jiarong. On the high optical spectrum interference imaging spectrometer[D](2008).

    [14] Xu G Q, Zhang Y F, Wan J W et al. Application of high-resolution three-dimensional imaging lidar[J]. Acta Optica Sinica, 41, 1628002(2021).

    Tools

    Get Citation

    Copy Citation Text

    Xue-Rong XU, Yi-Tian PENG, Ming-Jian GU, Teng-teng JIANG. Analysis of the effect of vibration on the homogeneity of the kinematic mechanism of an angular mirror translational interferometer[J]. Journal of Infrared and Millimeter Waves, 2025, 44(3): 425

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Infrared Optoelectronic System and Application Technology

    Received: Jan. 25, 2024

    Accepted: --

    Published Online: Jul. 9, 2025

    The Author Email: Yi-Tian PENG (yitianpeng@dhu.edu.cn), Ming-Jian GU (gumingj@sin.com)

    DOI:10.11972/j.issn.1001-9014.2025.03.012

    Topics