Small animals, especially mice and rats, are widely used in biomedical research. Various imaging techniques have been developed for small animal studies[
Chinese Optics Letters, Volume. 16, Issue 12, 121701(2018)
Dual-modal photoacoustic/CT imaging system
Photoacoustic (PA) tomography (PAT) breaks the barrier for high-resolution optical imaging in a strong light-scattering medium, having a great potential for both clinical implementation and small animal studies. However, many organs and tissues lack enough PA contrast or even hinder the propagation of PA waves. Therefore, it is challenging to interpret pure PAT images, especially three-dimensional (3D) PA images for deep tissues, without enough structural information. To overcome this limitation, in this study, we integrated PAT with X-ray computed tomography (CT) in a standalone system. PAT provides optical contrast and CT gives anatomical information. We performed agar, tissue phantom, and animal studies, and the results demonstrated that PAT/CT imaging systems can provide accurate spatial registration of important complementary contrasts.
Small animals, especially mice and rats, are widely used in biomedical research. Various imaging techniques have been developed for small animal studies[
PAT is a hybrid imaging technique that combines optical absorption contrast and ultrasonic detection[
In this study, we integrated PAT with an X-ray-based CT to develop a standalone system. Our results demonstrated that this dual-modality imaging system has the advantages of both PAT and CT. The dual-modal imaging system in this study contains a PAT and a micro-CT, as shown in Fig.
Sign up for Chinese Optics Letters TOC Get the latest issue of Advanced Photonics delivered right to you!Sign up now
Figure 1.PAT-CT imaging system. (a) A schematic top view of the system describing the main components. (b) A photograph of the system. The phantom is positioned in the center.
The PAT system contains a 48-element customized US linear array (by TomoWave. Inc., Houston, Texas, USA) to detect the PA signal. Each element has a size of
The micro-CT system consists of an X-ray tube (XRS-125-2-P-CW, Source-ray. Inc., USA) and a planar detector (Dexela 1512 CMOS X-ray Detector, Dexela Ltd., UK). The electrical voltage and the current of the tube were set to 80 kV and 0.5 mA, respectively. The anode is tungsten with a focal spot of 0.5 mm. To alleviate the beam-hardening artifact, a 4.5 mm thick aluminum sheet was placed in front of the X-ray tube. The X-ray detector has
We performed the imaging sequentially in two steps (the reason for this procedure will be explained later): CT imaging followed by a PAT. During the CT imaging, no water was filled in the water tank, and CT recordings were evenly acquired along 360 deg. Then we filled the water tank with deionized (DI) water and started the PAT imaging. The rotational stage moved continuously at a speed of 1.2 deg/s. The data acquisition was triggered by the laser at a repetition rate of 10 Hz. A total of 3000 PA data sets (each set stored
For PAT reconstruction, we used the image reconstruction algorithm described in Ref. [
Spatial calibration must be done to co-register PAT and CT images. In this study, we used a painted thin metal wire as the dual-modal contrast. After image reconstruction for both modalities, we loaded these two images into a 3D Slicer (version 4.6, slicer.org) and set the CT image of the wire in the center. Then we manually rotated and translated the corresponding PAT image to fuse them, finding the calibration parameters. Therefore, we can get the spatial rotational and translational parameters to be used for later image fusing.
The agar phantom was a cylinder made of 1% intralipid and 3% agar, and it had a diameter of 27 mm. There were three objects buried in the agar: one black carbon rod with a diameter of 0.5 mm, one helix, and one transparent glass tube, as shown in Fig.
Figure 2.Photographs of the two phantoms. (a) The agar phantom with the carbon rod, helix, and glass tube. The photo was taken after the experiment. (b) The chicken breast tissue phantom.
Figure 3.Agar phantom imaging results. (a) The 3D rendering result for the CT reconstruction. (b) The 3D rendering result for the photoacoustic imaging. (c) The co-registered rendering of the PAT and CT images (
To demonstrate the potential of the system for tissue study, chicken breast tissue was also used in the study. One carbon rod with a diameter of 0.5 mm, one metal wire with a diameter of 0.4 mm, and one thin glass tube were imbedded in chicken breast tissue. Part of the metal wire was painted black. We inserted the chicken tissue into a plastic bottle made of low-density polyethylene (LDPE), which has a near-acoustic impedance with water.
In order to remove air bubbles, the bottle was also filled with water, as shown in Fig.
Figure 4.Tissue phantom imaging results. (a) The 3D rendering result for the CT reconstruction. (b) The 3D rendering result for the photoacoustic reconstruction. (c) The co-registered image of the PAT and CT imaging (
Then we carried out the animal experiment on a 20 g BALB/c female nude mouse. The mouse was euthanatized by
Figure 5.Mouse imaging results. (a) The nude mouse picture that was taken before the experiment. (b) The 3D rendering result for the CT reconstruction. (c) The 3D rendering result for the photoacoustic reconstruction. (d) The co-registered image of the PAT and CT imaging (
In this study, we presented a standalone dual-modal PAT-CT imaging system. PAT provides the optical absorption contrast, and CT gives the mass density distribution. The system obtains these two important complementary pieces of information with a highly accurate spatial registration. The imaging depth was demonstrated to be more than 2 cm in scattering biological tissues. The performance of this system has been demonstrated by two types of phantom studies and mouse experiments.
Because the X-ray source used in this study has a relative low power, the attenuated X-ray, after passing through the water in the tank, is too weak to be detected. Therefore, we first imaged the target by CT without filling water in the tank. A more powerful X-ray source would enable two modalities to image simultaneously in future. In addition, because it is limited by the element’s size, the geometry of the linear probe, and the bandwidth, our current PAT system has an imaging resolution of
[24] L. Li, L. Zhu, C. Ma, L. Lin, J. Yao, L. Wang, K. Maslov, R. Zhang, W. Chen, J. Shi. Nat. Biomed. Eng., 1, 0071(2017).
[25] G. Huang, S. Yang, Y. Yuan, D. Xing. Appl. Phys. Lett., 99, 125(2011).
[27] L. Jing, X. Liang, Z. Deng, S. Feng, X. Li, M. Huang, C. Li, Z. Dai. Biomater., 35, 5814(2014).
[28] G. Tian, X. Zhang, X. Zheng, W. Yin, L. Ruan, X. Liu, L. Zhou, L. Yan, S. Li, Z. Gu. Small, 10, 4160(2015).
Get Citation
Copy Citation Text
Kaiye Xia, Xiaohui Zhai, Zhaoheng Xie, Kun Zhou, Yutao Feng, Guangjie Zhang, Changhui Li, "Dual-modal photoacoustic/CT imaging system," Chin. Opt. Lett. 16, 121701 (2018)
Category: Medical optics and biotechnology
Received: Aug. 12, 2018
Accepted: Oct. 24, 2018
Published Online: Dec. 7, 2018
The Author Email: Changhui Li (chli@pku.edu.cn)