Chinese Optics Letters, Volume. 8, Issue 11, 1098(2010)
On-line measurement of temperature and water vapor in CH4/air premixed flame using near-infrared diode laser
We establish a single diode laser sensor system to obtain temperature and water concentration in CH4/air premixed flame. Line-of-sight properties are analyzed, but line-of-sight results are not path average values for temperature measurements. The measurements are performed on a flat burner based on scannedwavelength direct absorption spectroscopy using two adjacent water lines at 7153.75 and 7154.35 cm?1. Real-time results are acquired using a data acquisition card with a Labview data processing program. The standard uncertainties of the temperature and water concentration measurements are 2.3% and 5.1%, respectively.
Get Citation
Copy Citation Text
Bo Tao, Xisheng Ye, Zhiyun Hu, Lirong Zhang, Jingru Liu, "On-line measurement of temperature and water vapor in CH4/air premixed flame using near-infrared diode laser," Chin. Opt. Lett. 8, 1098 (2010)
Received: Apr. 9, 2010
Accepted: --
Published Online: Dec. 3, 2010
The Author Email: