Accuracy of single-shot autocorrelation measurements of petawatt laser pulses
At the Shen Guang II (SGII) Petawatt Laser Facility, measurements of large-energy, single-shot laser pulses sometimes feature asymmetric autocorrelation signals, causing uncertainty in the measurement of compressed pulses. This study presents a method for defining and describing the asymmetry of autocorrelation signals. We discuss two sources of asymmetry: the nonuniform distribution of the near field excited by a beam, and the rotation of autocorrelator arms from the cylinder lens. The pulsewidth of an asymmetric autocorrelation signal is shorter than its real width. After updating the autocorrelator, the single-shot autocorrelator for the SGII petawatt laser exhibits a measurement uncertainty of below 12.3%. Recommendations on reducing asymmetry in large-energy, single-shot autocorrelation are discussed.
Get Citation
Copy Citation Text
Xiaoping Ouyang, Jingui Ma, Lin Yang, Shunxing Tang, Chong Liu, Yonghua Peng, Liejia Qian, Baoqiang Zhu, Jianqiang Zhu, Zunqi Lin. Accuracy of single-shot autocorrelation measurements of petawatt laser pulses[J]. Collection Of theses on high power laser and plasma physics, 2012, 10(1): 3989