Chinese Journal of Lasers, Volume. 39, Issue s2, 208007(2012)

Error Analysis of Adjusting and Alignment for Multiband Spectrometer of One Meter Solar Telescope

Fu Yu1、*, Cheng Xiangming1, Yuan Shu1,2, and Xu Zhi1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    Based on the theory about the design of spectrometer and error analysis, a method about error analysis for system alignment is proposed. By using Zemax software, the alignment error on the image quality is analyzed. The result indicates that the tilt and tip of slit is less than 0.5°, then the coma involved can be depressed by adjusting the collimating mirror; the self-rotating of slit less than 10″ can make the slant of spectrum lines less than one pixel; By adjusting the position of CCD, ±10 mm the defocus of collimating mirror is compensated. Then, the observed result is given.

    Tools

    Get Citation

    Copy Citation Text

    Fu Yu, Cheng Xiangming, Yuan Shu, Xu Zhi. Error Analysis of Adjusting and Alignment for Multiband Spectrometer of One Meter Solar Telescope[J]. Chinese Journal of Lasers, 2012, 39(s2): 208007

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Measurement and metrology

    Received: Jul. 12, 2012

    Accepted: --

    Published Online: Dec. 31, 2012

    The Author Email: Yu Fu (fiona@ynao.ac.cn)

    DOI:10.3788/cjl201239.s208007