NUCLEAR TECHNIQUES, Volume. 47, Issue 7, 070202(2024)

Absolute measurement of absorbed dose to water for γ-ray of 60Co by water calorimetry

Qian YANG... Jie FAN* and Shimeng XU |Show fewer author(s)
Author Affiliations
  • National Institute of Measurement and Testing Technology, Chengdu 610021, China
  • show less

    Background

    Most of domestic radiation therapy doses can only be traced back to air kerma of 60Co γ-ray at present in China. The uncertainty of directly tracing radiation therapy dose to absorbed dose to water is much smaller compared to tracing radiation therapy dose to air kerma, which must be converted to absorbed dose to water.

    Purpose

    This study aims to solve the problem of traceability and transfer of 60Co γ-ray absorbed dose to water, as well as the problem that the dose measurement of radiotherapy cannot be directly traced through the absorbed dose to water, NIMTT (National Institute of Measurement and Testing Technology) has established an absolute measurement device of 60Co γ-ray absorbed dose to water.

    Methods

    An absolute measurement device of 60Co γ-ray absorbed dose to water was established by National Institute of Measurement and Testing Technology (NIMTT) using water calorimeterfor direct measurement of absorbed dose to water. The correction factors for water absorbed dose measurement were obtained by experimental and simulation methods. A comparison of laboratory reproducing results was made between NIMTT and NRC (National Research Council, Canada) using the method of transmitting standard to further verify the accuracy and consistency of the measurement of 60Co γ-ray absorbed dose to water. Two PTW ionization chambers were used as the transfer standard.

    Results

    The reproduced results of two laboratories for absolute measurement of 60Co γ-ray absorbed dose to water are consistent within the relative standard uncertainty of 0.71%, and the normalized error value of En is -0.45.

    Conclusions

    The comparison results have verified that NIMTT laboratory has the ability of traceability and transmission of absorbed dose to water. The results of this study provide a reference for the absolute measurement of absorbed dose to water for 60Co γ-ray.

    Keywords
    Tools

    Get Citation

    Copy Citation Text

    Qian YANG, Jie FAN, Shimeng XU. Absolute measurement of absorbed dose to water for γ-ray of 60Co by water calorimetry[J]. NUCLEAR TECHNIQUES, 2024, 47(7): 070202

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Apr. 9, 2024

    Accepted: --

    Published Online: Aug. 27, 2024

    The Author Email: FAN Jie (范杰)

    DOI:10.11889/j.0253-3219.2024.hjs.47.070202