Opto-Electronic Engineering, Volume. 35, Issue 6, 42(2008)

Wavelength Calibration for Ultraviolet Spectroradiometer and Analysis of its Uncertainty

WANG Jia-peng1,2、*, WANG Shu-rong1, XU Ling-di1,2, SONG Ke-fei1, LIN Xue-song1, HUANG Yu1,2, WANG Long-qi1,2, WANG Rui1,2, ZHANG Zhen-duo1,2, and LI Fu-tian1
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    Referring to limitation of the conventional mercury lamp spectral calibration in ultraviolet wave band, a spectral calibration device was built for ultraviolet spectroradiometer. The physical process and measurement chain of spectral calibration were studied and twelve ozone absorptive spectral lines were calibrated. The measurement uncertainty and combined standard uncertainty of influenced quantities of wavelength calibration were analyzed and the combined standard uncertainty was 0.026nm. Spectral bandwidths can reach 1.0nm according to the analysis of optical CAD and experiment results. The spectral calibration system does well in ultraviolet spectral calibration for spectroradiometer. The calibrated results satisfy the spectral precision for ozone retrieval.

    Tools

    Get Citation

    Copy Citation Text

    WANG Jia-peng, WANG Shu-rong, XU Ling-di, SONG Ke-fei, LIN Xue-song, HUANG Yu, WANG Long-qi, WANG Rui, ZHANG Zhen-duo, LI Fu-tian. Wavelength Calibration for Ultraviolet Spectroradiometer and Analysis of its Uncertainty[J]. Opto-Electronic Engineering, 2008, 35(6): 42

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Jul. 10, 2007

    Accepted: --

    Published Online: Mar. 1, 2010

    The Author Email: Jia-peng WANG (wjp1009@126.com)