High Power Laser and Particle Beams, Volume. 34, Issue 12, 124001(2022)
High-energy CT system with 10 lp/mm spatial resolution
Based on the photocathode S-band electron linac, a high-energy (9 MeV) micro-focus CT system named “Jingwei” was built. The focal-point is less than 100 μm, and the 7 h dose fluctuation is as low as 1.6%. An initial imaging experiment was also carried out. The focal spot of the dual-wire image quality meter clearly distinguished the 13D wire with a wire diameter of 0.05 mm, and the CT spatial resolution measured by a test card is better than 10 lp/mm. The electron beam energy of the system is also adjustable from 6 MeV to 18 MeV.
Get Citation
Copy Citation Text
Qinghua Liu, Jing Li, Lijun Shan, Dexin Xiao, Qing Pan, Yu Liu, Hanbin Wang, Dongcai Hu, Peng Zhang, Shoutao Li, Jianxin Wang, Demin Zhang, Longgang Yan, Xiaoli Zhang, Kongyin Gan, Chengxin Zhang, Peng Li, Xuming Shen, Wei Bo, Yunbin Chen, Xiaohui Li, Shuaihua Wang, Yong Yu, Hao Chen, Xiutai Hu, Guowu Ma, Kui Zhou, Zheng Zhou, Yuan Wang, Xingfan Yang, Dai Wu, Ming Li, Menxue Chen, Jinguang Hu, Jianheng Zhao, Guobin Fan. High-energy CT system with 10 lp/mm spatial resolution[J]. High Power Laser and Particle Beams, 2022, 34(12): 124001
Category: Particle Beams and Accelerator Technology?Research Letter
Received: Oct. 1, 2022
Accepted: --
Published Online: Nov. 10, 2022
The Author Email: Wang Hanbin (wanghanbin@caep.cn), Wu Dai (wudai04@163.com)