Chinese Journal of Lasers, Volume. 36, Issue 10, 2528(2009)
Development of a Spectrum Domain 3D Optical Coherence Tomography System
We developed a 3D spectrum domain optical coherence tomography (OCT) system based on fiber-based Michelson interferometer. The measurable depth of the OCT imaging is up to 3.4 mm for the detection part is of high speed spectrometer with 0.05 mm spectrum resolution. The signal-noise-ratio (SNR) and axial resolution of this OCT system are experimentally found to be 51 dB and 8.5 m,respectively. A 3D OCT image could be acquired in two second owing to the fast A-scan rate (up to 10×103 line/s). Using 3D OCT technology,the images of biology tissue such as retina are shown more intuitive and precise.
Get Citation
Copy Citation Text
Duan Lian, He Yonghong, Zhu Rui, Ma Hui, Guo Jihua. Development of a Spectrum Domain 3D Optical Coherence Tomography System[J]. Chinese Journal of Lasers, 2009, 36(10): 2528
Category:
Received: May. 9, 2009
Accepted: --
Published Online: Oct. 23, 2009
The Author Email: Lian Duan (laolian.duan@gmail.com)