Chinese Journal of Lasers, Volume. 41, Issue 12, 1208008(2014)
Accurate Measurement of the Refractive Index D-Glucose Solution at Various Concentrations at Different Temperatures
Refractive index of D-Glucose solution in many areas such as scientific research and industry has very important applications. It′s very important to measure the refractive index accurately. The beam splitter prism interferometry is used to measure refractive index of D-Glucose solution at different temperatures with different concentrations. The basic principle of this method and apparatus is introduced. The refractive index of a sample of known calcium chloride solution is measured by the system and contrasted with literature values to verify its reliability. The accuracy is up to 10-4. In order to research the characteristics of D-Glucose solution, diversifications of D-Glucose solution′s refractive index at different temperatures measured and plotted into graphs. Error analysis is conducted as well. The results can provide help for medicine and biology.
Get Citation
Copy Citation Text
Song Zheyi, Feng Guoying, Zhang Tao. Accurate Measurement of the Refractive Index D-Glucose Solution at Various Concentrations at Different Temperatures[J]. Chinese Journal of Lasers, 2014, 41(12): 1208008
Category: measurement and metrology
Received: Jun. 9, 2014
Accepted: --
Published Online: Oct. 8, 2014
The Author Email: Zheyi Song (315552001@qq.com)