Experiment Science and Technology, Volume. 22, Issue 2, 8(2024)

Creative Adjustment of the Flow Cell Sorter and the Application in the Purification of Naive T Cells

Yajuan WAN, Rui WANG, Song ZHANG*, Ying ZHOU, and Jiafu LONG
Author Affiliations
  • College of Life Sciences, Nankai University, Tianjin 300071, China
  • show less

    How to adjust the flow cell sorter for the sensitive cells is a concern of many practitioners. Using the FACSAria Fusion as an example, the article creatively combined the service interface with the user interface to adjust the lowest default sample pressure and then sorted the primary naive T cells according to the standard user’s manual with the 85μm nozzle. The testing quality of the purified naive T cells revealed that the procedure could lead to high purity products. Moreover, treg cells differentiated from the products could generate Foxp3 normally compared with the non-sorting control.

    Keywords
    Tools

    Get Citation

    Copy Citation Text

    Yajuan WAN, Rui WANG, Song ZHANG, Ying ZHOU, Jiafu LONG. Creative Adjustment of the Flow Cell Sorter and the Application in the Purification of Naive T Cells[J]. Experiment Science and Technology, 2024, 22(2): 8

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Feb. 22, 2023

    Accepted: --

    Published Online: May. 21, 2024

    The Author Email: Song ZHANG (张松)

    DOI:10.12179/1672-4550.20230079