Infrasound waves are widely present in nature, with frequencies below the threshold of human hearing (
Photonics Research, Volume. 12, Issue 5, 969(2024)
Miniaturized and highly sensitive fiber-optic Fabry–Perot sensor for mHz infrasound detection
Infrasound detection is important in natural disasters monitoring, military defense, underwater acoustic detection, and other domains. Fiber-optic Fabry–Perot (FP) acoustic sensors have the advantages of small structure size, long-distance detection, immunity to electromagnetic interference, and so on. The size of an FP sensor depends on the transducer diaphragm size and the back cavity volume. However, a small transducer diaphragm size means a low sensitivity. Moreover, a small back cavity volume will increase the low cut-off frequency of the sensor. Hence, it is difficult for fiber-optic FP infrasound sensors to simultaneously achieve miniaturization, high sensitivity, and extremely low detectable frequency. In this work, we proposed and demonstrated a miniaturized and highly sensitive fiber-optic FP sensor for mHz infrasound detection by exploiting a Cr-Ag-Au composite acoustic-optic transducer diaphragm and a MEMS technique-based spiral micro-flow hole. The use of the spiral micro-flow hole as the connecting hole greatly reduced the volume of the sensor and decreased the low-frequency limit, while the back cavity volume was not increased. Combined with the Cr-Ag-Au composite diaphragm, a detection sensitivity of
1. INTRODUCTION
Infrasound waves are widely present in nature, with frequencies below the threshold of human hearing (
Fiber-optic Fabry–Perot (FP) acoustic sensors, based on the optical FP interference principle, have the advantages of high sensitivity, small structure size, long-distance detection, immunity to electromagnetic interference, easy to multiplexing networking, and so on [19–27]. For an FP acoustic sensor, acoustic-optic transducer diaphragms of various materials were usually considered to improve the detection sensitivity of acoustic pressure response, such as metallic diaphragms [28–31], silicon-related diaphragms [32–34], polymer diaphragms [35,36], two-dimensional materials diaphragms [37–40], and special structure diaphragms [41–44]. On the other hand, the detection sensitivity is directly proportional to the size of the diaphragm. When pursuing miniaturized sensors, the detection sensitivity will be sacrificed due to the small diaphragm size. Furthermore, the detection sensitivity of the sensor is not only constrained by the materials and sizes of transducer diaphragms but also depends on the back cavity structures, which are designed to improve the response characteristics and balance the influence of the environmental factors such as temperature. For infrasound detection, to decrease the low cut-off frequency, the back cavity volume can be increased, but this will also increase the size of the sensor. Another optional approach is mechanical filtering [45,46]. In 2021, Zhang
In this work, a fiber-optic FP sensor with a very small size of
Sign up for Photonics Research TOC Get the latest issue of Advanced Photonics delivered right to you!Sign up now
2. PRINCIPLES AND SIMULATION
A schematic diagram of the proposed fiber-optic FP acoustic sensor is shown in Fig. 1(a). The end surface of the fiber and the transducer diaphragm formed an FP interference resonator (FP cavity), which can be simplified as a two-beam interferometer due to the low reflectivity of the fiber end surface. When the acoustic wave acts on the sensitive transducer diaphragm, the diaphragm is elastically deformed, and thus the reflection interference spectrum is modulated by the acoustic wave.
Figure 1.(a) Schematic diagram of the fiber-optic FP acoustic sensor. (b) Electrical-mechanical-acoustic equivalent model of the sensor.
When the thin diaphragm vibrates, the edge experiences minimal displacement while the center experiences maximum since the surrounding of the diaphragm is clamped. For an FP sensor, the fiber end surface must align with the diaphragm center to determine the displacement sensitivity of the diaphragm by calculating the ratio of the shape variable at the midpoint to the applied acoustic pressure. The diaphragm’s tension is
When selecting diaphragm materials, theoretical analysis and calculations were conducted. On the other hand, for silicon-related diaphragms or polymer diaphragms, due to their low reflectivity, a layer of metal needs to be coated on them to improve reflectivity, which undoubtedly increases the production cost and process difficulty. As for two-dimensional materials, such as graphene, the large-scale preparation and the consistency of sensors are difficult to guarantee. By the calculation and simulation of diaphragm vibration theory, the Cr-Ag-Au composite metal diaphragm was selected as the acoustic-optic transducer diaphragm, which has high sensitivity and stable performance. Cr was selected as the bottom layer because Cr has excellent corrosion resistance and wear resistance, and Cr was easier to grow on the photoresist during evaporation. The temperature of Ag can be kept at a constant level at a low temperature during the coating process, which can control the residual thermal stress of the diaphragm caused by temperature change. As a result, the aggregation of thermal stress can be reduced, and thus the sensitivity can be further improved. Consequently, Ag was selected as the main deformation layer. However, Ag is prone to oxidation in the air to form silver oxide, which can affect the stability of the sensor performance. Hence, a layer of Au diaphragm was evaporated on the upper layer due to the good chemical stability of Au. In addition, Au can provide a high reflectivity for optical interferometry measurement. Furthermore, Ag is a face-centered cubic crystal structure and Cr is a body-centered cubic crystal structure. The interface composed of the two different crystal structures has a strong ability to hinder dislocation motion [48].
Moreover, to improve the low-frequency response of the FP acoustic sensor, the electrical-mechanical-acoustic equivalent model was applied to the sensor design, as shown in Fig. 1(b). In the studied acoustic system, the size of each element is much smaller than the wavelength of the infrasound wave, which can be analyzed as a lumped system [49]. For a diaphragm-based acoustic sensor, the diaphragm’s performance can be easily affected by external environments such as temperature if the sensor cavity is closed, so an open cavity structure can eliminate the influence of external factors by making a connecting hole in the sensor cavity. However, the low-frequency acoustic pressure response of the sensor will be reduced due to the existence of the connecting hole. In the equivalent circuit model, the actual response of the acoustic sensor is determined by the sensitivity
Figures 2(a) and 2(b) show the simulation results for the effect of connecting holes size on the transfer function. The presence of the connecting hole does not affect the sensor response in the flat area. However, in the low-frequency range, both the length and radius of the connecting hole greatly influence the frequency response of the sensor. Specifically, shorter connecting hole length and larger connecting hole radius can result in more severe low-frequency attenuation. In addition, the effects of back cavity volume on the transfer function and low cut-off frequency are shown in Figs. 2(c) and 2(d), respectively. It can be seen that a large back cavity volume can lower the cut-off frequency, but a large back cavity volume means a large sensor size. Hence, to enhance the sensor’s response sensitivity in the low-frequency band and minimize attenuation, it is recommended to increase the length of the connecting hole and decrease its radius as much as possible within the constraints of the structure and process, while ensuring that the sensor size does not increase.
Figure 2.(a) Effects of connecting hole radius on transfer function when connecting hole length is fixed. (b) Effects of connecting hole length on transfer function when connecting hole radius is fixed. (c) Effects of back cavity volume on transfer function. (d) Effects of back cavity volume on low cut-off frequency.
3. EXPERIMENTAL SETUP AND RESULTS
A. Device Design
The MEMS technique was applied to the processing of the precision structures of the composite diaphragm. As shown in Fig. 3(a), the photoresist was used as the sacrificial layer in a silicon wafer and the electron beam evaporation process was adopted to complete the preparation of the high-performance composite metal diaphragms. The thickness was regulated to 20 nm, 500 nm, and 30 nm of Cr, Ag, and Au layers, respectively, by controlling the process parameters, and the density of the composite diaphragm was
Figure 3.(a) Fabrication process of the diaphragm. (b) Fabrication process of the spiral micro-flow hole and assembly of the sensor. (c) Left: picture of the single layer of silicon chip before bonding. Right: scanning electron microscope (SEM) image of the spiral groove structure. (d) Left: picture of the dual layers of large-area silicon wafers after bonding. Right: picture of the fiber-optic FP sensor.
To decrease the low cut-off frequency of the FP infrasound sensor, a design of a spiral micro-flow hole was proposed based on the MEMS technique, which can effectively increase the connecting hole length and reduce the sensor size, as shown in Figs. 3(b) and 3(c). First, the photoresist as a mask was coated on two silicon wafers. Second, the designed patterns of two mirror-symmetrical miniature grooves with a same diameter of about 48 μm were transferred to the photoresist in the two silicon wafers by photoetching, respectively. Third, the two grooves were etched on the two silicon wafers, respectively. The spiral shape extended the total length of each groove to 96.6 mm. Then the photoresist was cleaned off. Fourth, the starting point of one spiral groove and the end point of another spiral groove were etched into two through-holes by using the deep silicon etching method, respectively. The through-holes for optical fiber penetration were simultaneously etched out. Finally, through the silicon–silicon bonding process for the two silicon wafers, the spiral micro-flow holes were formed by the mirror-symmetrical surface-etched spiral grooves inside two layers of silicon wafers. Due to the limitation of the silicon–silicon bonding process, two large-area silicon wafers were used in the above steps. Each small silicon chip with a size of
B. Experimental Testing
According to the principle of acoustic metrology, we constructed an aeroacoustic testing system and conducted a performance test on the fiber-optic FP sensor, as shown in Fig. 4. Since the coupling cavity can suppress the low-frequency noise in the environment, we utilized the coupling cavity comparison method to measure the sensor response in the low-frequency band (0.01–100 Hz). Moreover, the free-field comparison method was employed to measure the sensor response in the high-frequency band (100 Hz to 5 kHz) due to the uneven distribution of acoustic field in the coupling cavity.
Figure 4.Infrasound sensor calibration experimental system.
A C-band amplified spontaneous emission (ASE) light source module was implemented as the probe light source, with a wavelength range of 1526–1563 nm. Moreover, the WLI Fourier phase demodulation method was employed for acoustic signal extraction [50,51]. The ASE light was coupled into the fiber-optic FP sensor through a variable optical attenuator (VOA) and a fiber-optic circulator, and the reflected light signals were imported to a fiber Bragg grating analyzer (FBGA). The spectral signal was collected in real-time and input into the computer via the FBGA, which has a high sampling rate of 20 kHz to detect acoustic signals in a wider frequency range. The spectral signals were then processed by a LabVIEW program that integrated a WLI demodulation algorithm to obtain the phase change information.
A low-frequency coupling cavity was used as an acoustic source for the sensor’s low-frequency response measurement from 0.01 to 100 Hz. A commercial B&K 4193-L-004 microphone was exploited as the reference. Both the FP infrasound sensor and the reference microphone were simultaneously positioned at the two input ports of the low-frequency coupling cavity. To measure the high-frequency response from 100 Hz to 5 kHz, a commercial B&K 4190-L-001 microphone and a commercial speaker were used as the reference and the acoustic source, respectively. The FP infrasound sensor and the reference microphone were symmetrically placed on either side of the acoustic source axis at the free field with a distance of approximately 2.0 m from the test acoustic source. To reduce the influence of the environment, all experimental devices were placed in a fully anechoic chamber. When testing at both high frequency and low frequency, the signal generator of a B&K 3160-A-042 acoustic analyzer produced a sinusoidal signal and drove the test sound source to generate acoustic waves through a power amplifier with a gain of 18 dB. The output electrical signal of the reference microphone was collected and analyzed by the acoustic analyzer to obtain the value of test acoustic pressure. Furthermore, the detection sensitivity of the FP infrasound sensor can be obtained by the WLI Fourier phase demodulation algorithm. The photos of the sensor calibration system are shown in Appendix C.
C. Experimental Result
Figures 5(a)–5(c) show the measured frequency spectra of the sensor output signals by the WLI Fourier phase demodulation algorithm under the acoustic waves at 0.01 Hz, 0.05 Hz, and 0.1 Hz, respectively, which demonstrate the superior frequency resolution of the sensor. The demodulated 0.1 Hz acoustic signal with an excellent sine fitting, as shown in Fig. 5(d), means that the acoustic signal was accurately restored. In addition, the linear response of the sensor was also measured, as shown in Fig. 5(e). The linearity of the sensor reached 0.9999 at 5 Hz acoustic wave.
Figure 5.Frequency spectra of the sensor output signals under the acoustic waves with different frequencies: (a) 0.01 Hz, (b) 0.05 Hz, (c) 0.1 Hz. (d) Sinusoidal fitting of time domain signal of 0.1 Hz. (e) Linearity of the sensor at 5 Hz acoustic wave.
The measured acoustic pressure sensitivity of the sensor from 0.01 Hz to 5 kHz is shown in Fig. 6. The results indicate that the designed fiber-optic FP acoustic sensor can achieve 0.01 Hz acoustic detection and has a flat response from 0.01 to 2500 Hz with a sensitivity fluctuation of
Figure 6.Acoustic pressure sensitivity of the fiber-optic FP acoustic sensor from 0.01 Hz to 5 kHz.
The MDP is another critical index for an acoustic sensor, which is determined by both sensitivity and noise. To accurately assess the sensor noise characteristic, the sensor was placed in a fully anechoic chamber to effectively suppress the potential influence of environmental noise. The noise output signal, denoted as
Figure 7.Noise test results. (a) Noise spectrum, (b) minimum detectable pressure (MDP) under different frequencies.
4. CONCLUSION AND DISCUSSION
In conclusion, a fiber-optic FP infrasound sensor was designed and fabricated from principle to verification, which used a highly sensitive Cr-Ag-Au composite diaphragm as the acoustic-optic transducer diaphragm and a miniaturized MEMS spiral micro-flow hole with a total length of 96.6 mm as a connecting hole. The infrasound sensor achieved 0.01 Hz infrasound detection and a flat response from 0.01 to 2500 Hz with a sensitivity fluctuation of
Acknowledgment
Acknowledgment. We thank engineer Jun Su in the Center of Optoelectronic Micro & Nano Fabrication and Characterizing Facility, Wuhan National Laboratory for Optoelectronics of Huazhong University of Science and Technology for the support in test and analysis.
APPENDIX A: PRINCIPLES OF THE DIAPHRAGM VIBRATION
Theoretical analysis of the forced vibration of the diaphragm is needed to select the appropriate material for the energy conversion diaphragm. According to the characteristics of the forced vibration of the diaphragm, the vibration equation can be obtained:
In the formula,
The formula above explains that the amount of displacement on a diaphragm when acoustic pressure is applied depends on its distance from the center. The edge experiences minimal displacement while the center experiences maximum. For FP sensors, the fiber end face must align with the center to determine the displacement sensitivity of the diaphragm by calculating the ratio of the shape variable at the midpoint to the acoustic pressure applied. The diaphragm’s tension is
The simulation of the influence of various parameters on the diaphragm sensitivity is shown in Fig.
Figure 8.Relationship between diaphragm sensitivity and various parameters. (a) Stress, (b) radius, (c) thickness, (d) density.
APPENDIX B: PRINCIPLES OF THE ELECTRICAL-MECHANICAL-ACOUSTIC EQUIVALENT MODEL
When studying acoustic sensors, the parameters in acoustics can be equivalent to those in circuits, and it is possible to analyze the acoustic phenomenon by solving the equivalent circuit equation. But it should be noted that the electric acoustic equivalence model is mainly applicable to lumped systems, that is, the sizes of the acoustic objects (cavities, stubs, small holes, etc.) are much smaller than the wavelength of the acoustic wave. In the studied acoustic system, the size of each element is much smaller than the wavelength of the acoustic wave, which can be analyzed as a lumped system. For a diaphragm acoustic sensor, an open cavity structure with connecting holes can be used to eliminate sensor performance changes caused by external environments. However, the low-frequency response of the sensor will be reduced due to the connecting hole. In the equivalent circuit model, the actual response of the acoustic sensor is determined by the sensitivity
The internal structure and the equivalent circuit system of the proposed FP acoustic sensor are shown in Figs.
The lumped element definitions are shown in Table
Lumped Element Definitions
Model Parameter | Symbol | Definition |
---|---|---|
Diaphragm acoustic mass | ||
Diaphragm acoustic compliance | ||
Diaphragm radiation mass | ||
Diaphragm radiation resistance | ||
Hole acoustic mass | ||
Hole radiation resistance | ||
Cavity acoustic capacity |
It can be seen from the transfer function expression that the sensing system has two resonance frequencies:
APPENDIX C: ACOUSTIC TEST PHOTOS OF THE FIBER-OPTIC FP SENSOR
The photos of the sensor calibration system are shown in Fig.
Figure 9.Acoustic test photos of the fiber-optic FP sensor.
[6] M. A. Garces. Infrasonic signals generated by volcanic eruptions. IEEE International Symposium on Geoscience and Remote Sensing, 1189-1191(2000).
Get Citation
Copy Citation Text
Peijie Wang, Yufeng Pan, Jiangshan Zhang, Jie Zhai, Deming Liu, Ping Lu, "Miniaturized and highly sensitive fiber-optic Fabry–Perot sensor for mHz infrasound detection," Photonics Res. 12, 969 (2024)
Category: Optical Devices
Received: Jan. 25, 2024
Accepted: Mar. 2, 2024
Published Online: May. 6, 2024
The Author Email: Ping Lu (pluriver@mail.hust.edu.cn)
CSTR:32188.14.PRJ.519946