Optical Technique, Volume. 48, Issue 3, 323(2022)

Research on the testing method of the corneal ablation depth of excimer laser corneal ametropia cure system

HOU Rui, ZHAO Zhihui, and ZHAO Yue
Author Affiliations
  • [in Chinese]
  • show less

    In order to test the ablating accuracy and stability of excimer laser corneal ametropia cure system in refractive surgery, a method based on depth measurement of PMMA plate is proposed. Firstly, the theoretical value of corneal ablation depth is analyzed. Then the reasons of the difference between the measured ablation depth and estimated ablation depth of different excimer laser machines are analyzed. Finally, the relationship between the estimated ablation depth of cornea and the measured depth of PMMA plate is fitted and the fitting line is taken as the theoretical value. The experimental results show that by measuring the depth of PMMA plate pre-operation in operation mode and comparing it with the theoretical value, the error of the indicated value is within ±5%.

    Tools

    Get Citation

    Copy Citation Text

    HOU Rui, ZHAO Zhihui, ZHAO Yue. Research on the testing method of the corneal ablation depth of excimer laser corneal ametropia cure system[J]. Optical Technique, 2022, 48(3): 323

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Nov. 25, 2021

    Accepted: --

    Published Online: Jan. 20, 2023

    The Author Email:

    DOI:

    Topics