NUCLEAR TECHNIQUES, Volume. 47, Issue 3, 030603(2024)

Determination of primary metal elements in LiF-UF4 and LiF-ThF4 molten salts via inductively coupled plasma atomic emission spectrometry

Ling HAN1,2, Yuanyuan TANG1、*, Min GE1, Yuan QIAN1, and Hongtao LIU1
Author Affiliations
  • 1Shanghai Institute of Applied Physics, Chinese Academy of Sciences, Shanghai 201800, China
  • 2University of Chinese Academy of Sciences, Beijing 100049, China
  • show less

    Background

    LiF-UF4 and LiF-ThF4 are used as addition salts in thorium-based molten salt nuclear reactors. Monitoring the product quality during the preparation of molten salts necessitates the analysis of the contents of the main metal elements in LiF-UF4 and LiF-ThF4, namely, lithium, uranium, and thorium.

    Purpose

    This study aims to establish effective, rapid analysis methods for determination of primary metal elements in LiF-UF4 and LiF-ThF4 molten salts.

    Methods

    Firstly, the LiF-UF4 samples were treated with nitric acid and hydrogen peroxide whilst the LiF-ThF4 samples were dissolved in aluminum nitrate. Then, inductively coupled plasma atomic emission spectrometry was used as rapid analysis methods for measuring these two molten salts, and manganese was employed as the internal standard element to reduce the effect of signal drift.

    Results

    In the analytical method of LiF-UF4, the recoveries of Li and U are detected to be 99.6%~102.4% and 99.6%~101.8%, respectively. The relative standard deviations (RSDs) of Li and U are 0.2%~0.7%, and 1.1%~2.0%, respectively. In the analytical method of LiF-ThF4, the recoveries of Li and Th are detected to be 99.6%~102.3% and 99.6%~102.4%, respectively. The relative standard deviations of Li and Th are 1.9%~2.0%, and 0.3%~0.4%, respectively.

    Conclusions

    The proposed methods are simple, time-efficient, accurate, and suitable for rapid analyses of large numbers of samples.

    Keywords
    Tools

    Get Citation

    Copy Citation Text

    Ling HAN, Yuanyuan TANG, Min GE, Yuan QIAN, Hongtao LIU. Determination of primary metal elements in LiF-UF4 and LiF-ThF4 molten salts via inductively coupled plasma atomic emission spectrometry[J]. NUCLEAR TECHNIQUES, 2024, 47(3): 030603

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Sep. 1, 2023

    Accepted: --

    Published Online: Apr. 29, 2024

    The Author Email: TANG Yuanyuan (TANGYuanyuan)

    DOI:10.11889/j.0253-3219.2024.hjs.47.030603

    Topics