Opto-Electronic Engineering, Volume. 44, Issue 10, 966(2017)

Resolution improvement for SD-OCT system based on Zemax

Xuya Zhang, Yongkai Zhu*, Xinhua Zhang, and Mingcan He
Author Affiliations
  • [in Chinese]
  • show less

    The design and optimization for lens group in spectral domain optical coherence tomography (SD-OCT) system is fairly important to improve the imaging quality. The field lens in sample arm and the focus-ing lens in spectrometer are designed based on Zemax, and the imaging quality is evaluated according to spot diagram and wavefront map. The results show that the field lens can provide ideal imaging quality and good lat-eral resolution. And the focusing lens in the spectrometer makes the size of focal spot smaller than the pixel width of CCD to avoid the interference in each pixel of CCD. In this way, we observe considerable improvement in the resolution of the spectrometer.

    Tools

    Get Citation

    Copy Citation Text

    Xuya Zhang, Yongkai Zhu, Xinhua Zhang, Mingcan He. Resolution improvement for SD-OCT system based on Zemax[J]. Opto-Electronic Engineering, 2017, 44(10): 966

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Jun. 4, 2017

    Accepted: --

    Published Online: Nov. 27, 2017

    The Author Email: Zhu Yongkai (zykjiangsu@163.com)

    DOI:10.3969/j.issn.1003-501x.2017.10.004

    Topics