The China Sky Survey Telescope(CSST)is an optical broadband space observation instrument that co-orbits with the China Space Station,as shown in
Journal of Infrared and Millimeter Waves, Volume. 42, Issue 2, 188(2023)
Research on the adaptability of space environment about NbN superconducting SIS mixer for high sensitivity terahertz detection module
The high sensitivity terahertz detection module (HSTDM) is one of the scientific instruments of the China Sky Survey Telescope. HSTDM is a high-resolution spectrometer and the first space heterodyne receiver using niobium nitride (NbN)-based superconducting tunnel junction (Superconductor - Insulator - Superconductor (SIS)) mixer (the NbN SIS mixer). The NbN SIS mixer must meet the specification requested for a space environment, such as high operation reliability, robustness to vibration, cosmic irradiation, and thermal variation. This paper presents the space qualification tests performed on the NbN SIS mixer, including sine and random vibration tests, single-particle irradiation test, total dose radiation test, and thermal cycling test. The mixer’s performance analysis confirms that it can meet the space application requirements of HSTDM.
Introduction
|
The China Sky Survey Telescope(CSST)is an optical broadband space observation instrument that co-orbits with the China Space Station,as shown in
Figure 1.Schematic diagram of the co-orbit flight between CSST and China Space Station
High sensitivity terahertz detection module(HSTDM)is one of the scientific instruments onboard the CSST. HSTDM is a high-resolution spectrometer and the first space heterodyne receiver using NbN SIS mixers with nearly five times quantum-noise limited sensitivity. The specifications of HSTDM compared with HIFI band 1[
HSTDM is another device that uses SIS mixer receivers in space after Herschel HIFI and JEM/SMILES [
1 Design and characterization of NbN SIS mixer
The NbN SIS mixer chip is installed in a mixer block(as shown in
Figure 2.Images of the SIS chip and bias-T board in the block
We have characterized the performance of the NbN SIS mixer before the mechanical vibration and particle irradiation test.
Figure 3.I-V curves of NbN SIS mixer at room temperature and 4.2 K
Figure 4.LO-pumped I-V curve and IF power curves at 490 GHz
2 Vibration test
The size of the NbN SIS mixer chip is 2 000 μm×102 μm×51 μm,and the weight is about 27 μg,the tunnel junction area is about 1 μm2. The chip itself is not sensitive to mechanical vibration. The POGO connectorc and fixation of the mixer chip or bonding might fail after repeated thermal cycling test or due to the strong vibrations during launch. HSTDM can use POGO connectors because it has been proven on HIFI,but environmental tests still need to verify its reliability. Sticking and fixing tiny chips with hot-melt wax has been widely used in the packaging of superconducting mixers on ground-based telescopes,but HSTDM will apply this method in space for the first time,so it is necessary to verify the reliability of the mixer chips through mechanical tests. To prevent the bonding wires from breaking or the welding spots from falling off due to vibration during the launching process,wax was used to seal the bonding wires and the welding spots at both ends of the mixer chip to improve the mechanical resistance. Therefore we vibration-tested at room temperature,at the Technical Institute of Physics and Chemistry(TIPC),CAS,a complete demonstration model(DM)containing a real NbN SIS mixer. The tests include sine vibration and random vibration. The vibration specifications from the decomposition of CSST qualification vibration conditions in the three directions of X,Y,and Z are listed in
|
The mixer block was bolted on an interface plate and then installed on the vibration table,in different configurations to test all three axes X,Y,and Z. One accelerometer was fixed on the plate for monitoring the mechanical response of the mixer block during the vibration process.
First,there was no major resonance during the vibration test at low frequencies(<200 Hz). One small(1 g)resonance showed up near 1 260 Hz in the random test for the X axes,and one small(0.5 g)resonance showed up near 1 790 Hz in the random rest for the Z axes. It could be a resonance in the interface plate bolting,or in the mixer unit itself. Furthermore,such a high-frequency resonance presents no threat to the mixer.
Second,these tests were an opportunity to test the reliability of the SIS mixer mounting and gluing techniques,the POGO connectors,and the circuit boards soldered on the aluminum block. We found no damage on any mechanical part after the vibration tests,and the mixer chip is not loose.
Third,we tested the unpumped and LO-pumped I-V curves and the IF power for the hot and cold loads at 490 GHz to study the effect of mechanical vibration on the performance of NbN SIS mixer. The IF power outputs are basically the same before and after the vibration tests with the same pumping current of the mixer. Comparison of the NbN SIS mixer characteristics before and after the vibration test demonstrates no loss of function or performance degradation. The test results show that the fixation of the NbN SIS mixer chip and the bonding wire connection is reliable,and the POGO connector can adapt to the mechanical vibration environment.
3 Radiation test
CSST operates at Low-Earth Orbit(LEO)with an orbital altitude of about 400 km. The main sources of LEO charged particle radiation are:(1)Galactic cosmic rays(charged particles from outside the solar system);(2)Electrons and protons captured by the Earth's radiation belts;(3)Solar cosmic rays [
According to the influence of particle irradiation on the ESA-ISO detector,the probability that particles hit a 1 μm2 tunnel junction is about once every 107 seconds [
|
We performed 20 krad(Si)irradiation test on the NbN SIS mixer using Co60,while the irradiation time was 56 hours. The bias voltage and current of the NbN SIS mixer were changeless during the test.
According to the IF power corresponding to the cold and heat load,we can obtain the receiver noise temperature using the Y-factor method in the bias voltage range of 3.8-4.4 mV shown in
Figure 5.Receiver noise temperature before and after irradiation at 490 GHz
4 Thermal cycling and bake-out test
Thermal cycling test could evaluate and validate thermal design safety margins and accelerate exposure to potential defects in products,eliminate early failures,and improve product reliability. The thermal cycling test conditions of NbN SIS mixer are from 4 K to 300 K,with 20-30 cycles [
The internal structures of the HSTDM cryostat need to be baked at 80 ℃ before launching to remove the internal contamination gas. However,during junction fabrication,higher temperatures are used for baking the photoresist,and SIS mixer chips are heated at about 120 ℃ when mounted,so the bake-out test is OK for the NbN SIS mixer. We will test the mixer for its resistance to a 140-hour long,80 ℃ bake-out in a thermal vacuum chamber.
Through the performance test of the NbN SIS mixer after the thermal cycling test,the results show that the mixer can adapt to the working thermal environment.
5 Conclusion
This paper has described the tests space qualification of NbN SIS mixers,including mechanical vibration test,particle irradiation test,and thermal cycling test. The DM has successfully passed these tests. Properties characterization of the mixer before and after the tests demonstrates that the mixer and the integrated block can adapt space environment. The research results in this paper have guiding significance for the qualification model development of HSTDM.
[9] Manabe T. Development of Superconducting Submillimeter-Wave Limb-Emission Sounder(JEM/SMILES) aboard the International Space Station[J]. Journal of the Communications Research Laboratory, 49, 17(2002).
[11] Delorme Y, Salez M, Lecomte B et al. Space-qualified SIS mixers for Herschel Space Observatory’s HIFI Band 1 instrument[C].
[13] van de Stadt H, Gao J R, de Graauw T et al. Assessment of Submillimeter Heterodyne Mixers for Space applications[C](1996).
Get Citation
Copy Citation Text
Kun ZHANG, Ming YAO, Dong LIU, Bo-Liang LIU, Jing LI, Qi-Jun YAO, Sheng-Cai SHI. Research on the adaptability of space environment about NbN superconducting SIS mixer for high sensitivity terahertz detection module[J]. Journal of Infrared and Millimeter Waves, 2023, 42(2): 188
Category: Research Articles
Received: Aug. 29, 2022
Accepted: --
Published Online: Jul. 19, 2023
The Author Email: Sheng-Cai SHI (scshi@pmo.ac.cn)