Chinese Journal of Lasers, Volume. 8, Issue 3, 56(1981)
[in Chinese]
In this paper the principle, configuration, behaviour and error analysis of the bulk absorption calorimeter which has large measurement range for ps pulsed laser radiation are reported. The measured energy ranges from 2 mj to 20 j, the operating wavelength ranges from 0.3m to 2m and 4m to 10m, the cooling time constant is about 73~75sec., and the accuracy of measurement is 5%.
Get Citation
Copy Citation Text
[in Chinese]. [J]. Chinese Journal of Lasers, 1981, 8(3): 56
Category: laser devices and laser physics
Received: Jan. 4, 1980
Accepted: --
Published Online: Aug. 17, 2012
The Author Email:
CSTR:32186.14.