Optical Technique, Volume. 48, Issue 6, 710(2022)

Analysis and optimization of a medical X-ray monochromator

LUAN Mei1, ZHANG Jian1, MOU Kuanhou1, and LI Yao2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    Aiming at the application requirements of monochromatic X-ray source in the field of medical detection, an optimized design scheme of tunable X-ray fluorescence monochromator is proposed. Firstly, the penetration depth of X-ray photons in human skin, bones and fats under different X-ray energy bands were analysised, and the optimal target angle, X-ray generation efficiency and radiation characteristics were calculated by using the Monte Carlo N Particle Transport Code (MCNP), and the metal anode target material was optimized too. On this basis, the single-energy X-ray conversion efficiency of the secondary copper target under different anode voltages (kV) was calculated by establishing the target structure, and the proportion of the Kα characteristic spectrum was also calculated. These relevant works could provide a basis for the optimization and application of the medical X-ray fluorescence monochromator.

    Tools

    Get Citation

    Copy Citation Text

    LUAN Mei, ZHANG Jian, MOU Kuanhou, LI Yao. Analysis and optimization of a medical X-ray monochromator[J]. Optical Technique, 2022, 48(6): 710

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: May. 1, 2022

    Accepted: --

    Published Online: Jan. 20, 2023

    The Author Email:

    DOI:

    Topics