Chinese Journal of Lasers, Volume. 30, Issue 6, 551(2003)

Refractive Index Measurement with Fabry-Perot Interference Wavelength

[in Chinese]* and [in Chinese]
Author Affiliations
  • [in Chinese]
  • show less

    A technique for accurate measurement of the refractive indexes of liquid and gas was proposed. The approach, based on Fabry Perot (F P) interferometer and using the central wavelength of transmission spectrum to elicit the measurand, essentially eliminates the errors aroused from light source fluctuation. Such a system was developed. The optical signal was transmitted through fibers, and a tunable optical filter is used to collect the sensing data. Using the test data of refractive indexes of a set of alcohol solution with different concentrations, the analytical relation between refractive index and central wavelength of transmission spectrum was educed with least squares linear regression method, the regression error is less than 0 00005. Refractive indexes of distilled water and acetone were measured respectively; the measuring error is no more than 0 0001.

    Tools

    Get Citation

    Copy Citation Text

    [in Chinese], [in Chinese]. Refractive Index Measurement with Fabry-Perot Interference Wavelength[J]. Chinese Journal of Lasers, 2003, 30(6): 551

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: measurement and metrology

    Received: Nov. 7, 2002

    Accepted: --

    Published Online: Jun. 27, 2006

    The Author Email: (rxwei@mail.whut.edu.cn)

    DOI:

    Topics