Chinese Journal of Lasers, Volume. 8, Issue 6, 56(1981)
[in Chinese]
A method is recommended to measure the spectral distribution of pulsed light sources with a prism monochromator. It is found that when a GW-5A monochromator is used we have to insure identical spectral width of monochromatic light (FWHP) over the whole range of wavelength measured. For this, formulas and data are provided and typical results are plotted.
Get Citation
Copy Citation Text
[in Chinese]. [J]. Chinese Journal of Lasers, 1981, 8(6): 56