Chinese Optics Letters, Volume. 12, Issue s1, S11203(2014)
Thermally Induced Errors Modeling and Control for PCMA
We develop the kinematic model of portable coordinate measuring arm (PCMA) based on the research on its mechanical characteristics and working principle, and present a comprehensive thermally-induced error correction model for PCMA. Based on the model and nominal data provided by a highly accurate coordinate measuring machine, the technique for thermally-induced error compensation of PCMA is presented along with measuring a standard artifact at the corresponding temperature field. Experimental results demonstrate that the deviation of PCMA at different temperatures is controlled in the range of ±10 μm using this technique to calibrate the PCMA.
Get Citation
Copy Citation Text
Wanli Liu, Zhankui Wang, Xiaoyang Li, "Thermally Induced Errors Modeling and Control for PCMA," Chin. Opt. Lett. 12, S11203 (2014)
Category: Instrumentation, measurement, and metrology
Received: Aug. 20, 2013
Accepted: Oct. 19, 2013
Published Online: Jul. 31, 2014
The Author Email: