Infrared and Laser Engineering, Volume. 45, Issue 7, 705005(2016)

Laser source used in accurate measurement of mid-infrared absolute spectral responsivity of detectors

Zhao Kun1、*, Shi Xueshun1,2, Liu Changming1, Liu Yulong1, Chen Haidong1, Liu Hongbo1, Chen Kunfeng1, and Li Ligong1,2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    In order to measure the absolute spectral responsivity of mid-infrared detectors with high accuracy, the improvement of power stability and beam quality of mid-infrared laser was theoretically and experimentally demonstrated. The power stability of mid-infrared laser was improved to be better than 0.1% by using acousto-optic modulator and feedback-control electronics. Based on propogation properties of Gaussion beam, the parameters were calculated and experimental setup of spatial filter was constructed. Owing to the spatial filter, the beam quality of mid-infrared laser was improved. A stable and reliable laser source is provided for the measurement of absolute spectral responsivity of mid-infrared detectors.

    Tools

    Get Citation

    Copy Citation Text

    Zhao Kun, Shi Xueshun, Liu Changming, Liu Yulong, Chen Haidong, Liu Hongbo, Chen Kunfeng, Li Ligong. Laser source used in accurate measurement of mid-infrared absolute spectral responsivity of detectors[J]. Infrared and Laser Engineering, 2016, 45(7): 705005

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: 激光技术及应用

    Received: Nov. 5, 2015

    Accepted: Dec. 3, 2015

    Published Online: Aug. 18, 2016

    The Author Email: Kun Zhao (kunzhao41@163.com)

    DOI:10.3788/irla201645.0705005

    Topics