Acta Photonica Sinica, Volume. 49, Issue 3, 0330002(2020)

Detection Method of Dissolved Oxygen Concentration in Water Based on Single Source Frequency Domain Fluorescence Lifetime

Peng CHEN1... Zhi ZHAO2, Dong-dong ZHAO1,*, Yang-yang HAN2 and Rong-hua LIANG1 |Show fewer author(s)
Author Affiliations
  • 1College of Computer Science and Technology, Zhejiang University of Technology, Hangzhou 310023, China
  • 2College of Information and Engineering, Zhejiang University of Technology, Hangzhou 310023, China
  • show less
    Figures & Tables(13)
    Design of detecting dissolved oxygen device based on frequency-domain fluorescence lifetime
    Structure of optical path
    Flowchart for detecting dissolved oxygen by the means of single source frequency-domain fluorescence lifetime
    Detection principle of single light source phase modulation
    Fabrication of artificial fingers and fingerprints film
    Calibration chart of the proposed method
    Fluorescence lifetime trend with temperature increase
    Comparison chart of instrument data
    • Table 1. Results of dissolved oxygen concentration

      View table
      View in Article

      Table 1. Results of dissolved oxygen concentration

      O2:N2/%[O2]/(mg·L-1)θ1/(°)Δτ/μsτdo
      21:798.963-69.12533.1075.152
      10:904.296-71.18027.1934.944
      5:952.062-73.82019.6982.901
      1:990.404-78.3886.9161.903
      0.01:99.990.033-80.5320.9561.200
      0:1000.008-80.87601.024
    • Table 2. Error analysis

      View table
      View in Article

      Table 2. Error analysis

      [O2]/(mg·L-1)τdoTest results/(mg·L-1)Deviation/(mg·L-1)
      8.9634.9448.960-0.003
      4.2962.9014.3090.013
      2.0621.9032.039-0.023
      0.4041.2000.4380.034
      0.0331.0240.0370.004
      0.0081.000-0.017-0.025
    • Table 3. Segmentation analysis of experimental results

      View table
      View in Article

      Table 3. Segmentation analysis of experimental results

      [O2]/(mg·L-1)Linear relationship|Δ|max/(mg·L-1)Error radio/%
      0~5[O2]=2.276×τdo-2.292≤|0.03|0.6
      5~9[O2]=2.276×τdo-2.292≤|0.04|0.5
    • Table 4. Analysis of stability and effectiveness

      View table
      View in Article

      Table 4. Analysis of stability and effectiveness

      Detection methodMeasurements/(mg·L-1)Average value/(mg·L-1)Standard deviation/(mg·L-1)Relative standard deviation/%
      12345
      DOP1 sensor4.2504.2424.2544.2334.2634.2480.0110.27
      Proposed method4.2244.2264.2354.2424.2334.2320.0070.18
    • Table 5. Analysis of stability and effectiveness

      View table
      View in Article

      Table 5. Analysis of stability and effectiveness

      Detection methodThe dissolved oxygen concentration to be detected AB/s
      0%→21%(90%)21%→0%(90%)0%→21%(95%)21%→0%(95%)
      DOP1 sensor35364345
      Proposed method21262337
    Tools

    Get Citation

    Copy Citation Text

    Peng CHEN, Zhi ZHAO, Dong-dong ZHAO, Yang-yang HAN, Rong-hua LIANG. Detection Method of Dissolved Oxygen Concentration in Water Based on Single Source Frequency Domain Fluorescence Lifetime[J]. Acta Photonica Sinica, 2020, 49(3): 0330002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Spectroscopy

    Received: Nov. 23, 2019

    Accepted: Jan. 13, 2020

    Published Online: Apr. 24, 2020

    The Author Email: ZHAO Dong-dong (zhaodd@zjut.edu.cn)

    DOI:10.3788/gzxb20204903.0330002