Acta Photonica Sinica, Volume. 45, Issue 6, 604001(2016)

Measurement and Correction of Linearity of Photon Counters

HU You-bo1,2、*, LI Jian-jun2, XIA Mao-peng2, GAO Dong-yang2, and ZHENG Xiao-bing2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    Linear measurement device was set up based on the linear relationship between pump power and photons in spontaneous parametric down conversion process. The nonlinearity errors mechanism, linear condition and correction model of photon counting detector were analyzed. For the sake of realizing the linear measurement of the photon counting, the pump power was adjusted to change incident photon rate, and the experimental measurements were corrected. The results show that the measurement precision can be improved by measurement and correction. The maximum linear range rate of the incident photon could be expanded from 105/s to 107/s result from lower non-linear errors.

    Tools

    Get Citation

    Copy Citation Text

    HU You-bo, LI Jian-jun, XIA Mao-peng, GAO Dong-yang, ZHENG Xiao-bing. Measurement and Correction of Linearity of Photon Counters[J]. Acta Photonica Sinica, 2016, 45(6): 604001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Dec. 4, 2015

    Accepted: --

    Published Online: Jul. 26, 2016

    The Author Email: You-bo HU (hgdhyb2010@163.com)

    DOI:10.3788/gzxb20164506.0604001

    Topics