Acta Photonica Sinica, Volume. 40, Issue 11, 1686(2011)

Calibration of Phaseshifted Sensitivity of Optical Fiber Hydrophone in the Frequency Range 20 Hz to 10 kHz

CHEN Yi1、*, ZHANG Jun1, ZHANG Min2, and WANG Liwei2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    Through accurately measuring the phaseshifted sensitivity of the head of optical fiber hydrophone using phased generated carrier demodulation technique, calibration of phaseshifted sensitivity of optical fiber hydrophone in the frequency range 20 Hz to 10 kHz was realized. At the frequencies from 20 Hz to 1.25 kHz, the calibration method was standing wave tube comparison method. And at the frequencies from 1.25 kHz to 10 kHz, the calibration method was freefield toneburst comparison method. The No.35 TMD optical fiber hydrophone was calibrated by the calibration system. The calibration results show that the phasedshifted sensitivities measured by two methods are very close, and the deviation value of sensitivity at the frequency 1.25 kHz is only 0.8 dB. Analysis of measurement uncertainties prove that the expand uncertainty (at k=2) of this calibration system is 0.9 dB.

    Tools

    Get Citation

    Copy Citation Text

    CHEN Yi, ZHANG Jun, ZHANG Min, WANG Liwei. Calibration of Phaseshifted Sensitivity of Optical Fiber Hydrophone in the Frequency Range 20 Hz to 10 kHz[J]. Acta Photonica Sinica, 2011, 40(11): 1686

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: May. 26, 2011

    Accepted: --

    Published Online: Dec. 12, 2011

    The Author Email: Yi CHEN (y.chen@163.com)

    DOI:10.3788/gzxb20114011.1686

    Topics