Spectroscopy and Spectral Analysis, Volume. 42, Issue 2, 654(2022)

Simulation and Experimental Measurement of Cryogenic Radiometer Cavity With Different Structures

Hong-bo LIU*, Xue-shun SHI, Xin-gang ZHUANG, Peng-ju ZHANG, Chang-ming LIU, and Heng-fei WANG
Author Affiliations
  • The 41st Institute of China Electronic Technology Group Corporation, Qingdao 266555, China
  • show less
    Figures & Tables(9)
    The structure of cavity (1#, 2#, 3#, 4#)
    Schematic diagram of the cavity absorption experimental setup
    Distribution diagram of blackbody cavity irradiance (1#, 2#)
    Measurment(a): Stability of the laser power (70 min); (b): Stability of white standard signal US;(c): Stability of the cavity signal Uc; (d): Stability of the background signal UB
    Comparison of simulation results and measurement results of the cavity with different structures
    • Table 1. The cavity parameters

      View table
      View in Article

      Table 1. The cavity parameters

      No.Long/mmThick/mmWeight/gDiameter/mmInlet-Diameter/mmAngle
      1#750.121010Cone-30°
      2#750.121010Inclined-30°
      3#750.121010Inclined-30°
      4#750.12.2128Inclined-30°
    • Table 2. The absorptance of simulation parameter setup

      View table
      View in Article

      Table 2. The absorptance of simulation parameter setup

      No.ItemParameter
      1Wave-length632.8 nm
      2Laser Power1 W
      3Light diameter2 mm
      4追迹光线数量119 401
      5追迹门槛设定10-7
      6涂层材料吸收率0.94
      7涂层镜面反射系数0.05
      8朗伯漫反射系数(ABg)0.01
    • Table 3. The cavity parameters

      View table
      View in Article

      Table 3. The cavity parameters

      No.1#2#3#4#
      Absorptance0.999 6580.999 9520.999 9640.999 974
    • Table 4. Measurement results of cavity absorption coefficient

      View table
      View in Article

      Table 4. Measurement results of cavity absorption coefficient

      No.ItemWhite standardCavityBackgroundAbsorptance
      1#Signal(V̅)1.017 780.000 667 2790.000 217 8060.999 568±0.000 05
      Std(V)1.2×10-46.4×10-65.8×10-6
      2#Signal(V̅)1.041 720.000 315 9260.000 275 2750.999 962±0.000 005
      Std(V)9.2×10-57.5×10-64.9×10-6
      3#Signal(V̅)1.041 890.000 329 3860.000 268 8820.999 943±0.000 005
      Std(V)1.6×10-47.2×10-64.8×10-6
      4#Signal(V̅)1.013 920.000 340 9010.000 280 2620.999 942±0.000 005
      Std(V)1.9×10-46.9×10-64.3×10-6
    Tools

    Get Citation

    Copy Citation Text

    Hong-bo LIU, Xue-shun SHI, Xin-gang ZHUANG, Peng-ju ZHANG, Chang-ming LIU, Heng-fei WANG. Simulation and Experimental Measurement of Cryogenic Radiometer Cavity With Different Structures[J]. Spectroscopy and Spectral Analysis, 2022, 42(2): 654

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Jan. 5, 2021

    Accepted: Feb. 24, 2021

    Published Online: Apr. 2, 2022

    The Author Email: LIU Hong-bo (lhbcetc@163.com)

    DOI:10.3964/j.issn.1000-0593(2022)02-0654-06

    Topics