Acta Optica Sinica, Volume. 40, Issue 20, 2011002(2020)

Calibration and Preprocessing Method for Amplitude Divided Simultaneous Imaging Polarimeter

Wenyu Cui1, Wenjuan Huang1,2, and Weining Yi1、*
Author Affiliations
  • 1Key Laboratory of Optical Calibration and Characterization, Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei, Anhui 230031, China
  • 2University of Science and Technology of China, Hefei, Anhui 230026, China
  • show less
    Figures & Tables(21)
    Schematic of three-beam simultaneous polarization imaging system
    Calculation results of polarization measurement errors caused by various error sources of beam splitting polarization imaging system. (a) Polarizer oriented angle error ak; (b) light divided ratio error bk; (c) transmissivity non-uniformity error ck; (d) radiation responsivity non-uniformity error di,j,k; (e) pixel offset ek
    Calculation results of polarization angle measurement errors caused by various error sources of beam splitting polarization imaging system. (a) Polarizer oriented angle error ak; (b) light divided ratio error bk; (c) transmissivity non-uniformity error ck; (d) radiation responsivity non-uniformity error di,j,k; (e) pixel offset ek
    Flow chart of pre-processing calibration for beam splitting simultaneous polarization imaging system
    Calibration system of CCD non-uniformity. (a) Schematic diagram; (b) physical graph
    Three-channel original image pixel value. (a) Image of 0° polarizer; (b) image of 45° polarizer; (c) image of 90° polarizer
    Three-channel image pixel value after uniformity correction. (a) Image of 0° polarizer; (b) image of 45° polarizer; (c) image of 90° polarizer
    Three-channel radiation response difference calibration system. (a) Schematic diagram; (b) physical graph
    Calibration system for polarizer oriented angle. (a) Schematic diagram; (b) physical graph
    Fitting curves of pixel values of different polarized angles. (a) 0° polarization channel; (b) 45° polarization channel; (c) 90° polarization channel
    Comparison of polarization degree measurement error before and after correction for amplitude divided imaging polarimeter. (a) Before correction; (b) after correction
    Comparison of polarization angle measurement error before and after correction for amplitude divided imaging polarimeter. (a) Before correction; (b) after correction
    Stokes parameter images after preprocessing. (a) Image of I parameter; (b) image of Q parameter; (c) image of U parameter; (d) image of polarization degree; (e) image of polarized angle
    Comparison between original image and polarization fusion image. (a) Original image; (b) polarization fusion image
    • Table 1. Polarization imaging system parameters

      View table

      Table 1. Polarization imaging system parameters

      ParameterValue
      Detection band /nm500-1000
      Pixel number1392 ×1040
      Pixel size /(μm×μm)6.45×6.45
      CCD size /(cm×cm)24×36
      Digitalizing bit8 or 12
      Max FPS /(frame/s)30
      Focal length /mm51.7
      Polarization angles /(°)0,45,90
      Camera FOV /(°)10×7.6
    • Table 2. Non-uniformity σ of three-channel original imageunit: %

      View table

      Table 2. Non-uniformity σ of three-channel original imageunit: %

      Lightσ1σ2σ3
      12.742.843.33
      22.572.762.94
      32.552.692.79
      42.422.712.72
      52.372.622.66
      62.382.622.62
      72.332.592.60
      82.392.602.58
      Average2.472.682.78
    • Table 3. Non-uniformity σ of three-channel image after uniformity correctionunit: %

      View table

      Table 3. Non-uniformity σ of three-channel image after uniformity correctionunit: %

      Lightσ1σ2σ3
      10.820.800.95
      20.460.440.53
      30.400.360.44
      40.330.320.38
      50.290.270.33
      60.300.280.34
      70.260.240.31
      80.220.190.26
      Average0.390.360.44
    • Table 4. Three-channel image average after radiation response uniformity correction

      View table

      Table 4. Three-channel image average after radiation response uniformity correction

      Light2468
      DN¯(0°)556.441138.651685.002249.34
      DN¯(45°)555.431136.241686.512249.34
      DN¯(90°)555.851133.841686.632252.58
    • Table 5. Relative difference ratio of average DN of 3 channel images before and after radiation responsivity uniformity correctionunit: %

      View table

      Table 5. Relative difference ratio of average DN of 3 channel images before and after radiation responsivity uniformity correctionunit: %

      Light2468Average
      ε2Original15.3416.4517.1417.2316.54
      Corrected-0.18-0.210.090.00-0.08
      ε3Original-19.59-18.27-17.32-17.04-18.06
      Corrected-0.11-0.420.100.14-0.07
    • Table 6. Measuring results of polarizer oriented angles

      View table

      Table 6. Measuring results of polarizer oriented angles

      Nominal value /(°)Measuring value /(°)Polarizer oriented angle error /(°)
      00.000.00
      4543.26-1.74
      9088.32-1.68
    • Table 7. Calibration values of system errors

      View table

      Table 7. Calibration values of system errors

      Errorθk /(°)ak/(°)εkσkek
      Before correction00.000.00000.02470.00
      45-1.740.16540.02682.01
      90-1.68-0.18060.02782.12
      After correction00.000.00000.00390.00
      45-1.74-0.00080.00360.22
      90-1.68-0.00070.00440.18
    Tools

    Get Citation

    Copy Citation Text

    Wenyu Cui, Wenjuan Huang, Weining Yi. Calibration and Preprocessing Method for Amplitude Divided Simultaneous Imaging Polarimeter[J]. Acta Optica Sinica, 2020, 40(20): 2011002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Imaging Systems

    Received: Apr. 30, 2020

    Accepted: Jul. 10, 2020

    Published Online: Sep. 30, 2020

    The Author Email: Yi Weining (yiwn@aiofm.ac.cn)

    DOI:10.3788/AOS202040.2011002

    Topics