Acta Optica Sinica, Volume. 38, Issue 4, 0430001(2018)

Normalization of Quantitative Analysis for Bacteria Multi-Wavelength Transmission Spectroscopy in Water

Yuxia Hu1,2,3, Nanjing Zhao1,3、*, Tingting Gan1,3, Jingbo Duan1,3, Deshuo Meng1,3, Jianguo Liu1,3, and Wenqing Liu1,3
Author Affiliations
  • 1 Key Laboratory of Environment Optics and Technology, Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei, Auhui 230031, China
  • 2 University of Science and Technology of China, Hefei, Auhui 230026, China
  • 3 Anhui Key Laboratory of Environment Optical Detection Technology, Hefei, Anhui 230031, China
  • show less
    Figures & Tables(7)
    Transmission spectra of Escherichia coli suspension with three concentrations
    (a) Measured and calculated optical density of spectrum 1 after particle concentration normalization; (b) concentration fitting curves of three different transmission spectra
    (a) Measured and calculated optical density of spectrum 1 after maximum normalization; (b) concentration fitting curves of three different transmission spectra
    (a) Measured and calculated optical density of spectrum 1 after integral normalization; (b) concentration fitting curves of three different transmission spectra
    (a) Measured and calculated optical density of spectrum 1 after average normalization; (b) concentration fitting curves of three different transmission spectra
    • Table 1. Structural parameters of Escherichia coli obtained with the inversion of different normalization transmission spectra

      View table

      Table 1. Structural parameters of Escherichia coli obtained with the inversion of different normalization transmission spectra

      SpectrumParticle concentration normalizationMaximum normalizationIntegral normalizationAverage normalization
      D1 /μmD2 /μmD1 /μmD2 /μmD1 /μmD2 /μmD1 /μmD2 /μm
      11.49000.05001.75710.05011.27920.05011.27150.0501
      21.39000.05001.43510.05011.46350.05011.46020.0501
      31.49000.05001.47990.05011.40860.05251.40860.0500
    • Table 2. Concentrations obtained with four normalization methods and serial dilution plate counting method

      View table

      Table 2. Concentrations obtained with four normalization methods and serial dilution plate counting method

      No.Concentration from plate counting /(107 mL-1)Particle concentration normalizationMaximum normalizationIntegral normalizationAverage normalization
      Concentration/(107 mL-1)Error /%Concentration/(107 mL-1)Error /%Concentration/(107 mL-1)Error /%Concentration/(107 mL-1)Error /%
      13.42002.366830.801.000870.743.35122.013.43000.29
      22.52003.071221.872.71757.842.52480.192.54320.92
      33.50002.776620.672.923016.493.54241.213.53100.89
      Average relative error /%24.4531.691.140.70
    Tools

    Get Citation

    Copy Citation Text

    Yuxia Hu, Nanjing Zhao, Tingting Gan, Jingbo Duan, Deshuo Meng, Jianguo Liu, Wenqing Liu. Normalization of Quantitative Analysis for Bacteria Multi-Wavelength Transmission Spectroscopy in Water[J]. Acta Optica Sinica, 2018, 38(4): 0430001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Spectroscopy

    Received: Oct. 17, 2017

    Accepted: --

    Published Online: Jul. 10, 2018

    The Author Email: Zhao Nanjing (njzhao@aiofm.ac.cn)

    DOI:10.3788/AOS201838.0430001

    Topics