Acta Physica Sinica, Volume. 69, Issue 14, 145202-1(2020)

Energy fluence measurement for pulse hard X-ray

Zhao-Feng Su*... Ding-Guo Lai, Meng-Tong Qiu, Qi-Fu Xu and Shu-Qing Ren |Show fewer author(s)
Author Affiliations
  • State Key Laboratory of Intense Pulsed Radiation Simulation and Effect, Northwest Institute of Nuclear Technology, Xi’an 710024, China
  • show less
    Figures & Tables(10)
    The abridged general view of the fluence measurement system for pulse hard X-ray.
    The block diagram of the calibration platform composition.
    The field diagram of detector sensitivity calibration.
    Experimental diagram of energy fluence measurement.
    Typical waveform of energy fluence measurement.
    • Table 1. The parameter comparison of several common high density scintillators.

      View table
      View in Article

      Table 1. The parameter comparison of several common high density scintillators.

      闪烁体最强发射波长/nm闪烁衰减时间/ns折射率密度/g·cm–3相对光输出/%
      Bi4Ge3O124803002.157.1312.0—14.0
      BSO4802/30/1002.066.802.0—5.0
      LSO:Ce420401.827.4075.0
      PbWO4480102.168.200.5—1.0
      GSO:Ce45031/651.906.7120.0
      ZnWO448050002.207.8726.0
      CdWO454050002.307.9040.0
      CaWO454050002.307.9040.0
    • Table 2. The transmissivity of the photon transmitting scintillators of different thicknesses.

      View table
      View in Article

      Table 2. The transmissivity of the photon transmitting scintillators of different thicknesses.

      Thicknesses of scintillators /mm0.0110.0015.0020.0025.0030.0035.00
      Transmissivity/%97.82001.99700.87500.44400.22700.12000.0695
    • Table 3. Sensitivity of the detector at different energies.

      View table
      View in Article

      Table 3. Sensitivity of the detector at different energies.

      靶材光谱仪电压/kV光谱仪电流/mAX射线能量/keV本底电流/nA输出电流/nA剂量率/mGy·s–1灵敏度/10–17 C·cm2·MeV–1
      Pb1002575.03.924.28.785.29
      W953059.22.211.93.256.79
      Dy753546.03.729.450.773.87
      Sn506025.33.926.755.773.13
    • Table 4. Measurement results of energy fluence in five successive experiments.

      View table
      View in Article

      Table 4. Measurement results of energy fluence in five successive experiments.

      发次二极管电压/kV二极管电流/kA剂量/rad能注量/mJ·cm–2
      1705311415.636.0
      2741315342.435.0
      3686323293.437.0
      4746327309.336.6
      5718305325.634.8
      平均值719.2 ± 25.0316.2 ± 8.9337.3 ± 47.435.9 ± 1.0
    • Table 5. Theoretial values of pulsed hard X-ray energy fluence.

      View table
      View in Article

      Table 5. Theoretial values of pulsed hard X-ray energy fluence.

      能量/keV15306090120150
      能谱归一化数值0.11740.22600.21100.14650.11040.0865
      质能吸收系数/cm2·kg–197904080143696969
      能注量/ mJ·cm–20.047610.219965.859068.434046.354754.97756
      能量/keV180210300400500600
      能谱归一化数值0.05330.03190.01030.00360.00210.0004
      质能吸收系数/cm2·kg–129.129.129.129.129.129.1
      能注量/ mJ·cm–27.275544.360811.417930.497080.292620.06403
    Tools

    Get Citation

    Copy Citation Text

    Zhao-Feng Su, Ding-Guo Lai, Meng-Tong Qiu, Qi-Fu Xu, Shu-Qing Ren. Energy fluence measurement for pulse hard X-ray[J]. Acta Physica Sinica, 2020, 69(14): 145202-1

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Nov. 7, 2019

    Accepted: --

    Published Online: Dec. 28, 2020

    The Author Email:

    DOI:10.7498/aps.69.20191700

    Topics