Spectroscopy and Spectral Analysis, Volume. 41, Issue 12, 3676(2021)

Research on the Relationship Between Modulation Depth and Center of High Order Harmonic in TDLAS Wavelength Modulation Method

Hao CHEN1、1; 2;, Yu JU3、3;, and Li HAN1、1;
Author Affiliations
  • 11. Institute of Electrical Engineering, Chinese Academy of Sciences, Beijing 100190, China
  • 33. Beijing Aerospace Yilian Science and Technology Development Company, Beijing 100176, China
  • show less
    Figures & Tables(8)
    Second harmonic center amplitude, fourth harmonic center amplitude and modulation depth function curve
    Simulation signals of the second and fourth harmonics under different air pressures
    Schematic diagram of experimental device
    Measured second and fourth harmonic signals at different air pressures
    Curves of measured second harmonic and fourth harmonic center amplitude
    Relative error curve of mean value of mp under different pressure conditions
    • Table 1. Theoretical and practical values of modulation depth m and harmonic center amplitude ratio under different air pressure

      View table
      View in Article

      Table 1. Theoretical and practical values of modulation depth m and harmonic center amplitude ratio under different air pressure

      Pressure/atmTheoretical/mMeasured/mRelative Error/%Theoretical/I4f/I2fMeasured/I4f/I2fRelative Error/%
      0.37.136 57.009 31.780.756 30.752 5-0.50
      0.54.281 94.218 01.490.629 40.625 1-0.68
      0.73.058 53.016 21.380.525 80.521 3-0.86
      0.92.378 82.347 01.340.441 40.436 8-1.04
      1.11.946 31.921 01.300.372 70.368 2-1.21
      1.31.646 91.625 81.280.316 60.312 4-1.33
      1.51.427 31.409 21.270.270 80.266 9-1.44
    • Table 2. Central amplitude, amplitude ratio and modulation depth m of the second and fourth harmonics under different air pressures

      View table
      View in Article

      Table 2. Central amplitude, amplitude ratio and modulation depth m of the second and fourth harmonics under different air pressures

      Pressure
      /kPa
      I4f
      (ν=ν0)
      I2f
      (ν=ν0)
      I4f/I2fm
      177.975.29921.370.081 70.622 6
      158.299.541 006.530.098 90.698 0
      140.4130.541 093.990.119 30.784 5
      119.6179.841 192.870.150 80.914 4
      100.5258.531 289.450.200 51.120 1
      79.0360.361 378.460.261 41.384 5
      69.6432.301 408.640.306 91.598 5
      61.1504.631 430.330.352 81.835 5
      49.9604.061 442.500.418 82.226 7
      40.2712.331 430.800.497 92.810 3
      30.2846.301 378.530.614 04.061 0
      20.0842.881 251.390.673 65.028 1
      10.4752.85948.680.793 68.631 0
    Tools

    Get Citation

    Copy Citation Text

    Hao CHEN, Yu JU, Li HAN. Research on the Relationship Between Modulation Depth and Center of High Order Harmonic in TDLAS Wavelength Modulation Method[J]. Spectroscopy and Spectral Analysis, 2021, 41(12): 3676

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Nov. 27, 2020

    Accepted: --

    Published Online: Dec. 17, 2021

    The Author Email:

    DOI:10.3964/j.issn.1000-0593(2021)12-3676-06

    Topics