Opto-Electronic Engineering, Volume. 37, Issue 2, 74(2010)

Beam Alignment for Diameter Measurement of the Single Crystal Silicon Sphere

KANG Yan-hui1,*... ZHU Ji-gui1, LUO Zhi-yong2 and YE Sheng-hua1 |Show fewer author(s)
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    According to the requirement of precision diameter measurement of the single crystal silicon sphere, the principle and characteristics of the measuring system were presented, and mathematical models were established for calculating the ellipticity of concentric interference rings with different beam incident angles. Furthermore, the diameter measuring errors of the fringe center were studied in different incident angle situations. It is shown that the error can be up to 6.6 nm when the original distance between inner face of the etalon plate and the adjacent silicon sphere surface equals 13.2 mm, and the incident angle equals 10-3 rad. A novel method was proposed for accurately adjusting incident angle of laser beam relative to the plate. The experimental results show that the incident angle can be less than 10-5 rad by applying this method.

    Tools

    Get Citation

    Copy Citation Text

    KANG Yan-hui, ZHU Ji-gui, LUO Zhi-yong, YE Sheng-hua. Beam Alignment for Diameter Measurement of the Single Crystal Silicon Sphere[J]. Opto-Electronic Engineering, 2010, 37(2): 74

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Jul. 28, 2009

    Accepted: --

    Published Online: May. 28, 2010

    The Author Email: Yan-hui KANG (yanhuikang@163.com)

    DOI:10.3969/j.issn.1003-501x.2010.02.014