Spectroscopy and Spectral Analysis, Volume. 42, Issue 9, 2870(2022)

Determination of Thallium in Blood of Occupational Exposed Population by Inductively Coupled Plasma Mass Spectrometry With High Matrix Introduction

Fei ZHANG1,1; 5;... Xia HUA2,2; *;, Jia-ying YUAN3,3;, Fan YOU1,1;, Ren-cai YE4,4;, Li DING4,4; and Jian-mei ZHAO4,4; |Show fewer author(s)
Author Affiliations
  • 11. Sichuan Center for Disease Control and Prevention, Chengdu 610041, China
  • 22. Sichuan Center for Food and Drug Evaluation, Inspection & Monitoring, Chengdu 610017, China;
  • 33. Chengdu University of Traditional Chinese Medicine, Chengdu 610032, China
  • 44. Liangshan Zhou Center for Disease Control and Prevention, Liangshan 615000, China
  • show less
    Figures & Tables(5)
    Mass spectrogram of standard series and internal standard (a); Standard curve equation of thallium (b)
    Variation of internal standard element signals
    • Table 1. Instrumental parameters for HMI-ICP-MS

      View table
      View in Article

      Table 1. Instrumental parameters for HMI-ICP-MS

      参数设定值
      RF power1 550 W
      Cool gas flow15.0 L·min-1
      Auxiliary gas flow0.90 L·min-1
      Plasma gas flow15 L·min-1
      Nebulizer pump speed0.1 r·min-1
      Carrier gas flow rate1.05 L·min-1
      Acquisition modeMass spectrogram
      Tuning modeNo Gas
      Plasma modeHMI
    • Table 2. Analytical results of thallium in blood under different treatments by HMI-ICP-MS (n=6)

      View table
      View in Article

      Table 2. Analytical results of thallium in blood under different treatments by HMI-ICP-MS (n=6)

      冻干全血质
      量控制样品
      方法测定值/(μg·L-1)精密度/%误差/%参考值范围/(μg·L-1)
      111.7, 10.9, 11.0, 11.4, 11.0, 12.86.2811.9
      211.0, 10.5, 11.5, 10.2, 10.7, 9.755.774.79
      L-2310.6, 10.8, 10.2, 9.91, 10.7, 10.63.263.5210.1(8.1~12.1)
      411.2, 11.2, 11.9, 10.9, 11.6, 10.83.7110.4
      510.3, 11.1, 11.3, 11.1, 10.7, 10.24.256.34
      132.7, 31.5, 34.4, 33.8, 30.9, 34.24.453.39
      234.4, 33.9, 31.1, 33.8, 34.7, 32.64.024.84
      L-3333.4, 31.9, 33.8, 31.8, 32.4, 32.52.472.5531.8(29.0~34.5)
      433.8, 32.9, 34.3, 34.5, 31.3, 32.23.804.12
      533.5, 34.5, 32.1, 33.1, 34.2, 33.22.564.89
    • Table 3. Recovery and precision of the method (n=7)

      View table
      View in Article

      Table 3. Recovery and precision of the method (n=7)

      本底
      /(μg·L-1)
      加标
      /(μg·L-1)
      测得值/(μg·L-1)回收率
      /%
      批内精密度
      /%
      批间精密度
      /%
      <0.422.502.63, 2.70, 2.60, 2.59, 2.55, 2.61, 2.49103.82.52-
      2.71, 2.45, 2.61, 2.44, 2.57, 2.38, 2.64101.7-4.77
      <0.4215.015.2, 15.7, 16.3, 15.3, 15.2, 15.6, 16.1104.22.81-
      15.4, 14.0, 14.2, 15.4, 14.9, 15.1, 13.992.7-4.44
      <0.4275.075.6, 74.8, 73.7, 72.9, 75.2, 72.1, 74.398.81.71-
      75.4, 76.5, 77.3, 72.9, 71.2, 74.6, 73.699.3-2.84
    Tools

    Get Citation

    Copy Citation Text

    Fei ZHANG, Xia HUA, Jia-ying YUAN, Fan YOU, Ren-cai YE, Li DING, Jian-mei ZHAO. Determination of Thallium in Blood of Occupational Exposed Population by Inductively Coupled Plasma Mass Spectrometry With High Matrix Introduction[J]. Spectroscopy and Spectral Analysis, 2022, 42(9): 2870

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Jul. 17, 2021

    Accepted: Oct. 25, 2021

    Published Online: Nov. 17, 2022

    The Author Email:

    DOI:10.3964/j.issn.1000-0593(2022)09-2870-05

    Topics