Infrared and Laser Engineering, Volume. 52, Issue 11, 20230136(2023)

High-sensitivity mid-infrared radiometer under vacuum and low temperature

Jingxu Sun, Qiang Fei, Hongbo Xie, Shuo Wang, and Xinwang Xie
Author Affiliations
  • Ji Hua Laboratory, Foshan 528200, China
  • show less
    Figures & Tables(12)
    Schematic diagram of the mid-infrared radiometer
    Photo of the mid-infrared radiometer
    (a) Optical design results of the mid-infrared radiometer; (b) Spot diagram; (c) Aberration; (d) Field curvature and distortion
    (a) Radial mounting and fixing structure of lens; (b) Axial mounting and fixing structure of lens
    Schematic diagram of thermal control structure of modulator
    Gain measurement of the mid-infrared radiometer
    Block diagram of signal processing
    Calibration scene of the mid-infrared radiometer under vacuum and low temperature environment
    Calibration of temperature response of the mid-infrared radio-meter under low temperature condition
    Calibration of temperature response of the mid-infrared radiometer under high temperature condition
    • Table 1. Uncertainty analysis of radiation standards and transfer process of radiometer

      View table
      View in Article

      Table 1. Uncertainty analysis of radiation standards and transfer process of radiometer

      Source of uncertaintyValue
      Standard blackbody transmission0.41%
      Traceability uncertainty0.30%
      Uncertainty of blackbody emissivity0.10%
      Uncertainty of blackbody temperature0.15%
      Non-uniformity0.20%
      Non-lambertian0.10%
      Uncertainty of radiometer1.08%
      SNR1%
      Response nonlinearity0.3%
      Signal instability0.24%
      Environmental temperature effects0.1%
      Combined uncertainty1.16%
    • Table 2. Comparison of on-site calibration with National Institute of Metrology

      View table
      View in Article

      Table 2. Comparison of on-site calibration with National Institute of Metrology

      NumberSetting value/KMeasured value/KTrue value/K
      1240.0239.6239.5
      2300.0299.5299.7
      3400.0399.1399.2
      4500.0498.5498.7
    Tools

    Get Citation

    Copy Citation Text

    Jingxu Sun, Qiang Fei, Hongbo Xie, Shuo Wang, Xinwang Xie. High-sensitivity mid-infrared radiometer under vacuum and low temperature[J]. Infrared and Laser Engineering, 2023, 52(11): 20230136

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Mar. 13, 2023

    Accepted: --

    Published Online: Jan. 8, 2024

    The Author Email:

    DOI:10.3788/IRLA20230136

    Topics