Infrared and Laser Engineering, Volume. 52, Issue 10, 20230033(2023)

Research on spectral emissivity characteristics of blackbody coatings

Jingwei Ding1,2, Xiaopeng Hao2, Kun Yu1, Jian Song2, and Jingjing Zhou2
Author Affiliations
  • 1College of Physics, Henan Normal University, Xinxiang 453007, China
  • 2National Institute of Metrology, Beijing 100029, China
  • show less
    Figures & Tables(15)
    Emissivity measuring device
    Coating structure
    Three kinds of coating physical picture
    Comparison of spectral emissivity between vacuum and atmosphere of three coatings
    Spectral emissivity of GR and Nextel 811-21 coatings of different thickness
    SEM comparison of Nextel 811-21 coatings. (a) and (b) are 1000 × and 200 × thin coatings; (c) and (d) are 1000 × and 200 × thicker coatings
    SEM comparison of GR coatings. (a) and (b) are 1000 × and 200 × thin coatings; (c) and (d) are 1000 × and 200 × thicker coatings
    Emissivity of three coatings within 30° measured angles
    Comparison of spectral emissivity of GR coatings before and after irradiation
    Comparison of spectral emissivity of Nextel 811-21 coatings before and after irradiation
    Comparison of spectral emissivity of Nextel 811- 21+MWCNT coatings before and after irradiation
    • Table 1. Nextel 811-21 and GR coating thickness, emissivity comparison table

      View table
      View in Article

      Table 1. Nextel 811-21 and GR coating thickness, emissivity comparison table

      Coating numberAverage thickness/µmEmissivity
      GR-154.90.9780
      GR-262.10.9784
      GR-379.50.9757
      GR-484.90.9751
      GR-5130.60.9713
      GR-6168.90.9698
      Nextel 811-21-168.60.9783
      Nextel 811-21-272.90.9787
      Nextel 811-21-397.40.9763
      Nextel 811-21-4175.90.9714
      Nextel 811-21-5191.70.9699
      Nextel 811-21-6233.40.9687
    • Table 2. Space irradiation comparison table

      View table
      View in Article

      Table 2. Space irradiation comparison table

      Particle energy/ keV Electron flux/ e·cm−2Proton flux/ p·cm−2In-orbit year/ a
      1003.6×10158.1×10143
      6.0×10151.3×10155
      9.6×10152.2×10158
    • Table 3. Comparison table of emissivity of retained sample coatings

      View table
      View in Article

      Table 3. Comparison table of emissivity of retained sample coatings

      Retained sample coatingFirst timeSecond timeVariation
      GR0.97350.97330.0002
      Nextel 811-210.97340.97370.0003
      Nextel 811-21+MWCNT0.97250.97220.0003
    • Table 4. Comparison table of emissivity of sample coating before and after irradiation

      View table
      View in Article

      Table 4. Comparison table of emissivity of sample coating before and after irradiation

      In-orbit year/aCoating numberUnirradiatedIrradiatedCoating color change
      3GR-10.97370.9737Grayness
      Nextel 811-21-10.97340.9739Grayness
      Nextel 811-21+MWCNT-10.97290.9720Grayness
      5GR-20.97360.9732Grayness
      Nextel 811-21-20.97390.9747Grayness
      Nextel 811-21+MWCNT-20.97290.9718Grayness
      8GR-30.97320.9723Grayness
      Nextel 811-21-30.97370.9754Grayness
      Nextel 811-21+MWCNT-30.97230.9702Grayness
    Tools

    Get Citation

    Copy Citation Text

    Jingwei Ding, Xiaopeng Hao, Kun Yu, Jian Song, Jingjing Zhou. Research on spectral emissivity characteristics of blackbody coatings[J]. Infrared and Laser Engineering, 2023, 52(10): 20230033

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Photoelectric measurement

    Received: Jan. 17, 2023

    Accepted: --

    Published Online: Nov. 21, 2023

    The Author Email:

    DOI:10.3788/IRLA20230033

    Topics