Acta Optica Sinica, Volume. 43, Issue 4, 0417001(2023)

Design of Optical System for Nasal Mucociliary Microendoscope

Xin Wang1、aff, Yang Xiang1、*, Dawei Feng1、aff, Kaiming Yang1、aff, Chuan Pang2,3、affaff, and Lei Chen2,3、affaff
Author Affiliations
  • 1School of Optoelectronic Engineering, Changchun University of Science and Technology, Changchun 130022, Jilin, China
  • 2Department of Otorhinolaryngology Head and Neck Surgery, Chinese PLA General Hospital, Beijing 100039, China
  • 3National Clinical Medical Research Center for Otorhinolaryngology, Beijing 100039, China
  • show less
    Figures & Tables(14)
    Structure diagrams of the prism viewing direction. (a) Front view of the prism viewing direction; (b) right view of the prism viewing direction
    Optical structure of endoscope objective lens
    Optical structure of a single set of rod mirrors
    Optical structure of eyepiece system
    Optical structure of multiplication adapter system. (a) Short focal; (b) medium focal; (c) long focal
    Schematic of microscopic objective
    Structure diagrams of equivalent microscopic objective. (a) Short focal; (b) medium focal; (c) long focal
    Overall optimization structure diagrams. (a) Combination of objective lens system and relay system; (b) rear short focal structure; (c) rear medium focal structure; (d) rear long focal structure
    MTF curves. (a) Short focal MTF; (b) medium focal MTF; (c) long focal MTF
    Spot diagrams. (a) Short focal spot diagram; (b) medium focal spot diagram; (c) long focal spot diagram
    Field curvature and distortion diagrams. (a) Short focal field curvature and distortion diagram; (b) medium focal field curvature and distortion diagram; (c) long focal field curvature and distortion diagram
    • Table 1. Design requirements of rigid endoscope objective

      View table

      Table 1. Design requirements of rigid endoscope objective

      System parameterValue
      Angle of view θ /(°)30
      Field of view 2y /mm0.40
      Square numerical aperture NA0.15
      Lens diameter C /mm<5.00
      Working distance d /mm3.00
      Wavelength λ /nm486-656
      Magnification2×
    • Table 2. Tolerance value

      View table

      Table 2. Tolerance value

      ToleranceRange
      Radius2 rings
      Thickness0.0375
      X direction surface parallel eccentricity0.01
      Y direction surface parallel eccentricity0.01
      X direction element parallel eccentricity0.01
      Y direction element parallel eccentricity0.01
      X direction surface tilt eccentricity0.016°
      Y direction surface tilt eccentricity0.016°
      X direction element tilt eccentricity0.016°
      Y direction element tilt eccentricity0.016°
      Refractive index0.001
      Abbe number0.1
      Degree of irregularity0.5
    • Table 3. Tolerance analysis results

      View table

      Table 3. Tolerance analysis results

      Probability /%Short focal MTF thresholdMedium focal MTF thresholdLong focal MTF threshold
      90≥0.24112795≥0.24939491≥0.25553497
      80≥0.26500962≥0.27334284≥0.27568140
      50≥0.31212522≥0.31387543≥0.31762968
      20≥0.34402264≥0.34406522≥0.34734586
      10≥0.35464655≥0.36532988≥0.36915205
    Tools

    Get Citation

    Copy Citation Text

    Xin Wang, Yang Xiang, Dawei Feng, Kaiming Yang, Chuan Pang, Lei Chen. Design of Optical System for Nasal Mucociliary Microendoscope[J]. Acta Optica Sinica, 2023, 43(4): 0417001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Medical optics and biotechnology

    Received: Aug. 15, 2022

    Accepted: Sep. 22, 2022

    Published Online: Feb. 16, 2023

    The Author Email: Xiang Yang (xyciom@163.com)

    DOI:10.3788/AOS221613

    Topics