Journal of Applied Optics, Volume. 44, Issue 4, 756(2023)

Simple medical rigid endoscopic relative percent distortion detection system

Lin LIU1... Yang XIANG1,*, Lixin LIU1, Weidong AN2 and Lei HUANG2 |Show fewer author(s)
Author Affiliations
  • 1Changchun University of Science and Technology, Changchun 130022, China
  • 2Jilin Institute of Metrology-Jilin Key Laboratory of metrology and testing instruments and technology, Changchun 130103, China
  • show less
    Figures & Tables(13)
    Schematic diagram of 30° laparoscope
    Relative percent distortion curve of endoscope
    Pattern of calibration target plate
    Schematic diagram of calibration
    Distortion of target plate
    Schematic diagram of overall system design
    Flow chart of system work
    Physical image of overall testing device
    Diagram of software interface
    Coincidence images of calibration target plate
    Calibrated relative percent distortion measurement image
    Relative percent distortion detection images without vertical calibration
    • Table 1. Measurement results of relative percent distortion of endoscope%

      View table
      View in Article

      Table 1. Measurement results of relative percent distortion of endoscope%

      次数方式方块位置平均值方差
      1校准24.124.824.323.224.10.6
      2校准24.424.324.523.924.30.2
      3校准23.824.424.323.624.00.3
      4未校准24.727.128.823.626.12.0
      5未校准27.224.524.129.826.42.3
      6未校准24.128.129.224.826.62.2
    Tools

    Get Citation

    Copy Citation Text

    Lin LIU, Yang XIANG, Lixin LIU, Weidong AN, Lei HUANG. Simple medical rigid endoscopic relative percent distortion detection system[J]. Journal of Applied Optics, 2023, 44(4): 756

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Oct. 27, 2022

    Accepted: --

    Published Online: Aug. 10, 2023

    The Author Email:

    DOI:10.5768/JAO202344.0401008