High Power Laser and Particle Beams, Volume. 32, Issue 3, 035005(2020)

Energy spectrum measurement for pulsed hard X-ray from 15 keV to 600 keV

Zhaofeng Su, Dingguo Lai, Mengtong Qiu, Shuqing Ren, Qifu Xu, and Shi Yang
Author Affiliations
  • State Key Laboratory of Intense Pulsed Radiation Simulation and Effect (Northwest Institute of Nuclear Technology), Xi’an 710024, China
  • show less
    Figures & Tables(7)
    Change of scatter-transmission ratio under difference distance of the detectors
    Collimating device of the detector
    Energy deposition curves of photon on the detectors after absorption plate of different mass thickness
    Photogragh of the energy spectrometer in the experiment
    Measured waveforms of the spectrometer
    Spectrum comparison between experiment energy and theoretical calculation
    • Table 1. Distance between the detector and the copper absorption plate when scatter-transmission ratio is 5%

      View table
      View in Article

      Table 1. Distance between the detector and the copper absorption plate when scatter-transmission ratio is 5%

      thickness of the copper absorption plate/mm1.002.164.105.017.007.9810.0712.00
      distance between the detector and the copper absorption plate/cm0.421.01.82.62.93.54.45.1
    Tools

    Get Citation

    Copy Citation Text

    Zhaofeng Su, Dingguo Lai, Mengtong Qiu, Shuqing Ren, Qifu Xu, Shi Yang. Energy spectrum measurement for pulsed hard X-ray from 15 keV to 600 keV[J]. High Power Laser and Particle Beams, 2020, 32(3): 035005

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Sep. 14, 2019

    Accepted: --

    Published Online: Mar. 19, 2020

    The Author Email:

    DOI:10.11884/HPLPB202032.190354