High Power Laser and Particle Beams, Volume. 34, Issue 2, 024001(2022)

Research of flying focal spot of the CT tube based on dynamic magnetic focusing

Wei Fu... Yanmei Wang, Xiangying Lü and Peng Wang |Show fewer author(s)
Author Affiliations
  • Beijing Vacuum Electronics Research Institute, Beijing 100015, China
  • show less
    Figures & Tables(11)
    Structural diagram of the electrostatic focusing system
    Dynamic magnetic focusing flying focus scanning structure of double magnetic quadrupole lens CT tube
    Magnetic field distribution of magnetic quadrupole lens
    CST simulation structure of dual magnetic quadrupole lens system
    Maximum magnetic field vs focal spot size in x and y axes
    Distribution of the magnetic field of quadrupole lens along the x-y axis
    Current simulation trajectory
    Morphology of focal spot without axial deflection current
    Relationship between x, y-axis of fset of focal spot and x, y-axis deflection current
    Add the x, y-axis deflection current deflection
    • Table 1. Application of foreign institutions in dynamic magnetic focusing flying focus technology

      View table
      View in Article

      Table 1. Application of foreign institutions in dynamic magnetic focusing flying focus technology

      numberinstitutionCT tubeflying focal spot technologyimage resolution improvement
      1Siemens[7]Stratonz-Sharp technology(z-axis double sampling)each probe unit yields two alternately overlapping X-line projections without a dose increase, substantially improving the spatial resolution and imaging quality
      2PhilipsiMRCprecise flying focal spot (SFS,the x-axial and z-axial flying focal spot technology)double sampling rate is obtained in plane and longitudinal direction, which is equivalent to 4 times of detector, and high spatial resolution is obtained in axial position and spiral scanning
      3DunleeCT5000xDFS(the x-axial flying focal spot),zDFS(the z-axial flying focal spot)improve imaging quality, provide sharp and clear images, zDFS double the number of slices, and there is no need for detectors with higher spatial resolution
    Tools

    Get Citation

    Copy Citation Text

    Wei Fu, Yanmei Wang, Xiangying Lü, Peng Wang. Research of flying focal spot of the CT tube based on dynamic magnetic focusing[J]. High Power Laser and Particle Beams, 2022, 34(2): 024001