High Power Laser and Particle Beams, Volume. 34, Issue 2, 024001(2022)

Research of flying focal spot of the CT tube based on dynamic magnetic focusing

Wei Fu... Yanmei Wang, Xiangying Lü and Peng Wang |Show fewer author(s)
Author Affiliations
  • Beijing Vacuum Electronics Research Institute, Beijing 100015, China
  • show less

    CT tube is the core component of CT equipment. CT tube adopts dynamic magnetic focusing technology to facilitate the realization of small focus when there is large current output. Flying focal spot technology can record each projection of each scan from multiple angles, improve the sampling rate, and greatly improve the image definition and quality. CST software is used to simulate the CT tube based on dynamic magnetic focusing. When the CT tube outputs high current of 1 A, the key parameters of the dual magnetic quadrupole lens are optimized to get the target size of focus of 0.2 mm×0.6 mm and the function of flying focus.

    Tools

    Get Citation

    Copy Citation Text

    Wei Fu, Yanmei Wang, Xiangying Lü, Peng Wang. Research of flying focal spot of the CT tube based on dynamic magnetic focusing[J]. High Power Laser and Particle Beams, 2022, 34(2): 024001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Particle Beams and Accelerator Technology

    Received: Sep. 2, 2021

    Accepted: --

    Published Online: Jan. 26, 2022

    The Author Email:

    DOI:10.11884/HPLPB202234.210392

    Topics