Laser & Optoelectronics Progress, Volume. 55, Issue 6, 061204(2018)
Calibration of System Parameters Based on Direct Phase Measuring Deflectometry
Phase measuring deflectometry (PMD) based on fringe reflection has been widely studied as a way of obtaining three-dimensional shape of specular objects. System calibration is an important step, and it determines the accuracy of the measurement results. We propose a calibration method to obtain the system parameters based on phase information. As a result, it can build the relationship between the absolute phase map and depth data. A contrast experiment is done for verification about extrinsic parameters of the LCD screen by phase data and the checkerboard. The experiment shows that the method using phase data is more accurate when images are out of focus. Using the calibration system, we test a concave mirror and an artificial specular step with discontinuous reflective surface, and the error is about 22 μm. Experiment results show that the proposed method can precisely determine the system parameters, so that 3D shape of specular objects can be measured with a high accuracy.
Get Citation
Copy Citation Text
Xiaoting Deng, Nan Gao, Zonghua Zhang. Calibration of System Parameters Based on Direct Phase Measuring Deflectometry[J]. Laser & Optoelectronics Progress, 2018, 55(6): 061204
Category: Instrumentation, Measurement and Metrology
Received: Oct. 8, 2017
Accepted: --
Published Online: Sep. 11, 2018
The Author Email: Zhang Zonghua (zhzhang@hebut.edu.cn)