Acta Photonica Sinica, Volume. 53, Issue 4, 0401003(2024)

Polarization Accuracy Verification and Out-of-band Response Analysis for Aviation Polarization Radiometer

Zhengdong XU1...2, Pingping YAO2,*, Mengfan LI2, Xiangjing WANG2, Yanghao CHU3, Yadong HU2, Jiexiang CHENG2 and Jin HONG2 |Show fewer author(s)
Author Affiliations
  • 1Institutes of Physical Science and Information Technology, Anhui University, Hefei 230061, China
  • 2Anhui Institute of Optics and Fine Mechanics, Hefei Institutes of Physical Science, Chinese Academy of Science, Hefei 230031, China
  • 3School of Instrument Science and Opto-electronics Engineering, Hefei University of Technology, Hefei 230009, China
  • show less
    Figures & Tables(13)
    Principle of aviation polarization radiometer system
    Schematic of the main structure of the airborne polarimetric radiometer
    Schematic of single-channel optical system for airborne polarimetric radiometer
    Airborne polarimeter radiometric calibration site
    Calibration of polarization measurement accuracy of airborne polarimetric radiometer
    Polarization measurement error is affected by out-of-band response and the degree of polarization
    Scene for the relative spectral response validation of a specific channel in the 490 nm wavelength band
    Relative spectral response of a specific channel in the 490 nm wavelength band
    The transmittance curve of filters used in the same batch
    • Table 1. Instrument parameter of the airborne polarimetric radiometer

      View table
      View in Article

      Table 1. Instrument parameter of the airborne polarimetric radiometer

      ParameterValue
      Spectral band

      550 nm, 910 nm

      490 nm ,670 nm,870 nm,1 610 nm

      Bandwidth20~60 nm
      Polarized angle0°, 60°, 120°
      Instantaneous field of view≥1.9°
      Radiometric calibration accuracy≤3%
      Polarization measurement accuracy<1%@0.2 degree of polarization
    • Table 2. Analysis results of absolute radiometric calibration uncertainty for airborne polarimetric radiometers

      View table
      View in Article

      Table 2. Analysis results of absolute radiometric calibration uncertainty for airborne polarimetric radiometers

      The spectral band of the APRua/%NLk/%NSk/%uk,a/%
      490P12.310.820.352.48
      490P22.310.810.342.47
      490P32.310.900.372.51
      5502.350.390.092.38
      670P12.240.200.062.25
      670P22.240.180.062.25
      670P32.240.160.062.25
      870P11.990.090.102.00
      870P21.990.090.102.00
      870P31.990.090.101.99
      9101.960.270.161.99
      1610P11.740.510.101.82
      1610P21.740.550.101.83
      1610P31.740.520.101.82
    • Table 3. The measurement accuracy of 490 channels polarization of the instrument was verified by comparative experiment

      View table
      View in Article

      Table 3. The measurement accuracy of 490 channels polarization of the instrument was verified by comparative experiment

      Light sourceRotation anglePCkPΤkΔPk
      0.000 00.000 00.000 0
      Tungsten-halogen lamp38.3°0.103 20.091 50.011 7
      51°0.207 50.184 60.022 9
      0.000 00.000 00.000 0
      LED monochrome light source38.3°0.103 20.103 10.000 1
      51°0.207 50.208 10.000 6
    • Table 4. Verification results of polarization measurement accuracy of airborne polarization radiometer

      View table
      View in Article

      Table 4. Verification results of polarization measurement accuracy of airborne polarization radiometer

      Band/nmRotation anglePCkPΤkΔPk
      0.000 00.000 00.000 0
      49038.3°0.103 20.103 10.000 1
      51°0.207 50.208 10.000 6
      0.000 00.000 00.000 0
      67038.3°0.101 30.102 10.000 8
      51°0.204 00.204 20.000 2
      0.000 00.000 00.000 0
      87038.3°0.103 20.098 70.001 6
      51°0.204 00.200 90.001 2
      0.000 00.000 00.000 0
      1 61038.3°0.097 80.097 70.000 1
      51°0.197 60.196 70.000 9
    Tools

    Get Citation

    Copy Citation Text

    Zhengdong XU, Pingping YAO, Mengfan LI, Xiangjing WANG, Yanghao CHU, Yadong HU, Jiexiang CHENG, Jin HONG. Polarization Accuracy Verification and Out-of-band Response Analysis for Aviation Polarization Radiometer[J]. Acta Photonica Sinica, 2024, 53(4): 0401003

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Oct. 13, 2023

    Accepted: Nov. 15, 2023

    Published Online: May. 15, 2024

    The Author Email: YAO Pingping (ppyao@aiofm.ac.cn)

    DOI:10.3788/gzxb20245304.0401003