Chinese Journal of Quantum Electronics, Volume. 41, Issue 2, 300(2024)

Design of high‑precision two‑dimensional calibration platform control system for detector array

YU Longkun1... LI Zhixiang2,3, WEN Xiaohui2,3, LIU Hui2,3 and TAN Fengfu3,* |Show fewer author(s)
Author Affiliations
  • 1School of Advanced Manufacturing, Nanchang University, Nanchang 330031, China
  • 2School of Information Engineering, Nanchang University, Nanchang 330031, China
  • 3Key Laboratory of Atmospheric Optics, Anhui Institute of Optics and Fine Mechanics, HFIPS,Chinese Academy of Sciences, Hefei 230031, China
  • show less
    Figures & Tables(13)
    Overall design structure diagram of two-dimmensionl calibration platform for detector array
    Schematic diagram of single point scanning track
    Schematic diagram of 3 unit × 3 unit detector array target module scanning track
    Schematic diagram of hardware circuit system
    Flow chart of lower computer software control system
    Schematic diagram of incremental PID closed loop control system
    Step response curve of incremental PID control stepping motor position
    Physical picture of two-dimensional calibration platform control system. (a) Front; (b) Back; (c) Equipment of repeated positioning accuracy experiment; (d) Operation picture
    Repeated positioning error of X-axis
    Repeated positioning error of Y-axis
    A curve of detector unit response value in single point mode test
    • Table 1. Results of ten single point tests

      View table
      View in Article

      Table 1. Results of ten single point tests

      Number of experimentsResponse maximumAverage valueStandard deviationCV
      11053610525.413.68860.0013
      210528
      310519
      410533
      510521
      610540
      710522
      810497
      910543
      1010515
    • Table 2. Processing results of plane calibration data of partial detector units

      View table
      View in Article

      Table 2. Processing results of plane calibration data of partial detector units

      No.Measure A valueBackground A valueCalibration coefficient /(Wˑcm-2ˑA-1)

      Measuring power density

      /(Wˑcm-2)

      Output power

      density

      /(Wˑcm-2)

      Normalized

      ratio

      Measurement

      error

      111812.513569.050.01410116.2329114.791.012570.01257
      210342.613331.710.01659116.3109114.791.013250.01325
      310188.333342.620.01719117.6781114.791.025160.02516
      49119.773375.320.02018115.9229114.791.009870.00987
      59359.753148.420.01893117.5581114.791.024120.02412
      610131.833346.950.01712116.1424114.791.011780.01178
      79260.723365.920.01972116.2455114.791.012680.01268
      811317.353281.410.01445116.0790114.791.011230.01123
      99273.513487.550.01948112.6822114.790.98164-0.01836
    Tools

    Get Citation

    Copy Citation Text

    Longkun YU, Zhixiang LI, Xiaohui WEN, Hui LIU, Fengfu TAN. Design of high‑precision two‑dimensional calibration platform control system for detector array[J]. Chinese Journal of Quantum Electronics, 2024, 41(2): 300

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Dec. 17, 2022

    Accepted: --

    Published Online: Jun. 24, 2024

    The Author Email: TAN Fengfu (fftan@aiofm.ac.cn)

    DOI:10.3969/j.issn.1007-5461.2024.02.012

    Topics