Acta Optica Sinica, Volume. 41, Issue 24, 2422003(2021)

Design of Compact Scleral Angiography Device for Eye Movement Tracking

Dong Xu and Weihua Song*
Author Affiliations
  • School of Instrument Science and Optoelectronic Engineering, Beihang University, Beijing 100191, China
  • show less
    Figures & Tables(18)
    Relative position of compact sclera angiography device
    Model of imaging device
    Optical path of compact sclera angiography device
    Structure of compact sclera angiography device. (a) Integrated optical device; (b) section view of imaging device
    Equivalent linear optical path diagram in vacuum
    Subgraph shape and distribution diagram
    Simulation diagram of image spatial resolution
    Simulation diagram of system field angle
    Simulation diagram of system depth of field
    Schematic of depth of field simulation for microlens section
    Main lens and microlens parameter setting
    Image of scleral vessels
    Optical system structure
    Simulation result
    Partial enlargement of eye-striae region in single subgraph
    Evaluation results of transversal image quality. (a) Modulation transfer function curves; (b) dot plot
    • Table 1. Parameters of imaging system

      View table

      Table 1. Parameters of imaging system

      ParameterSymbolValue
      Distance between object and main lensz1 /mm25
      Distance between main lens and microlensz2 /mm11.165
      Distance between microlens and image sensorz3 /mm1.163
      Size of apertureS/mm2.4
      Pixel sizep0 /mm0.002
      Sensor resolutionP /(pixel×pixel)2688×1520
      Microlens diameterdw /mm0.25
      Focal length of main lensfm /mm25
      Focal length of microlensfw /mm1.163
    • Table 2. Simulation parameter

      View table

      Table 2. Simulation parameter

      Surface typeRadiusThicknessGlassSemi-diameter
      Standard 1Infinity25.0009.000
      Paraxial5.400H-K91.600
      Coordinate 100
      Standard 2Infinity0Mirror5.000
      Coordinate 2-7.1420
      Coordinate 300
      Standard 3Infinity0Mirror5.000
      Coordinate 43.5000
      Standard 4Infinity0.900F_SILICA3.000
      Lens array-0.5351.1633.000
      Standard 5Infinity2.800
    Tools

    Get Citation

    Copy Citation Text

    Dong Xu, Weihua Song. Design of Compact Scleral Angiography Device for Eye Movement Tracking[J]. Acta Optica Sinica, 2021, 41(24): 2422003

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Optical Design and Fabrication

    Received: Apr. 19, 2021

    Accepted: Jun. 28, 2021

    Published Online: Nov. 30, 2021

    The Author Email: Song Weihua (397008531@qq.com)

    DOI:10.3788/AOS202141.2422003

    Topics