Acta Optica Sinica, Volume. 43, Issue 3, 0312006(2023)

Measurement Method of Inner Surface Size of Hole Based on Distortion Correction of Endoscopic Image

Qiang Sheng1,2、*, Jianming Zheng1、**, Ting Chen1, Xiuxiu Yang3, and Xianming Gao2
Author Affiliations
  • 1School of Mechanical and Precision Instrument Engineering, Xi′an University of Technology, Xi′an, 710048, Shaanxi, China
  • 2College of Mechanical and Electrical Engineering, Shaanxi University of Science and Technology, Xi′an, 710021, Shaanxi, China
  • 3School of Economics and Management, Xi′an University of Technology, Xi′an710054, Shaanxi, China
  • show less
    Figures & Tables(16)
    Endoscopic imaging principle. (a) Spatial coordinate system; (b) phase plane coordinate system
    Coordinate system of corrected endoscopic image
    Principle of axial correction. (a) Distortion effect; (b) correction process
    Measurement scheme and equipment. (a) Measurement scheme; (b) calibration broad; (c) endoscopic probe sleeve; (d) endoscopic image
    Preprocessing and center detection results. (a) Preprocessed result; (b) edge detection result; (c) center detection result
    Circumferential correction result
    Axial correction result
    MSE of network model
    3D surface fitting result
    Experimental process and shooting results. (a) Endoscopic image of Φ18 hole with trunking; (b) endoscope image of M18 thread hole; (c) Φ18 hole; (d) simulated defect; (e) endoscopic image of Φ18 hole
    Correction and edge detection results. (a) Correction result of experiment 1; (b) contour information of experiment 1; (c) correction result of experiment 2; (d) contour information of experiment 2; (e) defect local amplification of experiment 3; (f) correction result of experiment 3; (g) contour information of experiment 3
    • Table 1. Calibration block width statistics

      View table

      Table 1. Calibration block width statistics

      Line number123456789
      After circumferential correction37.3337.6737.8338.1738.1738.3338.3338.1737.67
    • Table 2. Calibration block height statistics

      View table

      Table 2. Calibration block height statistics

      Correction methodHeightMean valueStandard deviation
      Line number 1Line number 2Line number 3Line number 4Line number 5

      Line

      number

      6

      Line number 7Line number 8Line number 9
      Before axial correction21.3318.5016.0013.5011.8310.6710.179.479.1613.404.322
      After axial correction21.2520.9620.8120.3720.7520.9221.5721.3521.6721.070.421
    • Table 3. Experimental results of different apertures

      View table

      Table 3. Experimental results of different apertures

      ImageΦ12Φ16Φ21Φ25Φ28Φ32
      Original images
      Corrected image
    • Table 4. Pixel position and measurement error statistics

      View table

      Table 4. Pixel position and measurement error statistics

      ParameterVertex coordinate

      Edge length

      pixel distance /pixel

      Mean value /pixelStandard deviation

      Unit pixel size /

      (mm×mm)

      Measuring

      result /mm

      Real size /mmError /%
      Φ12

      (1569,46),(1910,287),

      (1776,678),(1373,684),

      (1248,274)

      665.04,667.82,

      664.83,667.43,

      662.13

      665.452.050.0182×0.018212.11120.92
      Φ16

      (1596,15),(1836,197),

      (1728,496),(1458,491),

      (1353,210)

      498.78,478.87,

      471.62,495.60,

      483.17

      485.6110.200.0250×0.025012.14121.17
      Φ21

      (1557,237),(1904,518),

      (1761,978),(1327,972),

      (1183,503)

      768.57,734.20,

      748.14,770.15,

      721.16

      748.4419.100.0328×0.032824.55251.80
      Φ25

      (1619,51),(1970,295),

      (1836,713),(1418,724),

      (1278,294)

      696.66,699.10,

      697.80,702.37,

      692.00

      697.593.390.0368×0.036825.67252.68
      Φ28

      (1614,23),(1927,208),

      (1822,579),(1451,582),

      (1322,216)

      593.63,605.35,

      617.87,582.28,

      605.05

      600.8412.040.0424×0.042425.48251.92
      Φ32

      (1653,19),(1914,205),

      (1817,524),(1502,530),

      (1398,227)

      530.96,524.76,

      513.59,532.84,

      516.47

      523.727.640.0462×0.046224.20253.20
    • Table 5. Pixel position and error statistics

      View table

      Table 5. Pixel position and error statistics

      ExperimentSample size

      Mean value

      of pixel

      distance /pixel

      Standard

      deviation of

      pixel distance

      Unit pixel size /

      (mm×mm)

      Measuring

      result /mm

      Real size /mmError /%
      Experiment 1268111.514.920.1095×0.109512.2113.147.08
      Experiment 247948.432.420.1095×0.10955.305.006.06
      Experiment 3560.405.460.0323×0.03231.952.107.10
    Tools

    Get Citation

    Copy Citation Text

    Qiang Sheng, Jianming Zheng, Ting Chen, Xiuxiu Yang, Xianming Gao. Measurement Method of Inner Surface Size of Hole Based on Distortion Correction of Endoscopic Image[J]. Acta Optica Sinica, 2023, 43(3): 0312006

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Jul. 20, 2022

    Accepted: Aug. 29, 2022

    Published Online: Feb. 13, 2023

    The Author Email: Sheng Qiang (zjm@xaut.edu.cn), Zheng Jianming (shengqiang@sust.edu.cn)

    DOI:10.3788/AOS221502

    Topics